16
PHCN-961
connection diagram
Warning
:
Make sure to power off before connecting. The back
panel consists of three connectors.
PHCN-961 series connection diagram
NEW VERSION
1. pH value 4—20mA output, positive terminal
14. Alternating connect earth wire
2. pH value 4—20mA output, negative terminal
15. Electric source input 110V alternating current
3. 485A output
16. Electric source input 220V alternating current
4. 485B output
17. Relay A (SP1)
5.
Y
Connector
REF
terminal
18. Relay A (SP1)
6. Short circuit with 7 under normal status
19. Relay B (SP2)
7. COM, open circuit with 6 pin and connect PMP sensor
when display value unstable
20. Relay B (SP2)
8. Temperature electrode positive terminal (PT1000)
21. Clearing Relay (RL3)
9. Temperature electrode negative terminal (PT1000)
22. Clearing Relay (RL3)
10. Temperature electrode sensing terminal (2-wire
system, use short circuit between 8 and 9)
__
pH/ORP input terminal
11. +5V
12. COM
13. Connect bigger earth wire
NOTE: Short circuit with 6 and 7 pin under normal status. Open circuit with 6 and 7 pin,
and connect PMP sensor when display value unstable.
17
SP1
SP2
12
N
13
14
15
16
11
10
09
08
07
06
05
04
03
02
01
mA
mA
pH-
COM
18
19
20
21
22
pH/ORP INPUT
REF
+5V
COM
NEW VERSION
485A
REF
485B
RL3