![Omega OSHFV Series Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/omega/oshfv-series/oshfv-series_user-manual_3957763006.webp)
(mm)
Mechanical Installation
The OS-HFV50 Model comes with a standard 3 m cable, a mounting nut and fixed mounting bracket. You
can mount the sensor in brackets or cutouts of your own design. For easy mounting and aligning the
sensor to the measured object a fixed or adjustable mounting bracket is available.
For exact measurement of the object temperature the sensor must be aligned correctly onto the object.
Mount the sensor so the measured spot is the same or smaller than the target.
Wiring
Position the sensor so the two laser beams are marking the exact location and size of the target.
Please gently move it around until the temperature signal reads the highest. Hold the sensor in
place and secure the mounting base. At ambient temperature > 50°C, the laser will be switched
off automatically.
LASER LIGHT
DO NOT STARE INTO BEAM
CLASS 2 LASER
<1mW / 650 nm
IEC 825 (1994)
red ------ 24VDC power
(
+
)
black ---- 24VDC power ( - )
white---- 4~20mA signal (+)
green --- 4~20mA signal ( - )
orange --- TX (TTL),optional
grey ----- RX (TTL), optional
blue ----- GND (TTL), optional
bare ----- Shield Ground
WARNING: Do not point the laser directly at the
eyes of persons or animals! Do not stare into the
laser beam. Avoid indirect exposure via reflective
surfaces!