background image

WHAT’S IN THE PACKAGE

The OS730K comes in a soft pouch inside a box along with a “K” type bead thermocouple

probe, a “9V” battery and this user’s manual.

PRODUCT OVERVIEW

Fig. 1 shows all controls, indicators, connectors and other physical features of the OS730K.

Fig. 2 shows all possible indications on the unit’s LCD. Familiarize yourself with the positions

of all controls and connectors and the meaning of all display indications and icons before

moving on to the Setup Instructions and Operating Instructions.

1. Laser pointer
2. Infrared sensor
3. Measurement trigger
4. Battery compartment
5. Jack for spade-type 

thermocouple probe

6. 

(laser/backlight) button

7. 

MODE

button

8. 

°C/°F

button

9. LCD

10. Laser identification/certification/

warning/safety labels 

(on left side of unit); 

see below

4

Fig. 1. All controls, indicators, connectors 

and other physical features of the OS730K

9

8

7

6

10

2

1

3

4

5

Содержание OS730K

Страница 1: ...OS730K 30 1 Ultra Wide Range IR Thermometer with K Port e mail info omega com For latest product manuals omegamanual info Shop online at omega comSM User sGuide MADE IN CHINA...

Страница 2: ...notice Servicing North America U S A Omega Engineering Inc One Omega Drive P O Box 4047 Stamford CT 06907 0047 USA Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada o...

Страница 3: ...30 1 ULTRA WIDE RANGE IR THERMOMETER WITH K PORT USER S MANUAL OS730K Please read this manual carefully and thoroughly before using this product...

Страница 4: ...using the thermometer The OS730K measures surface temperature using either of two methods non contact using an IR thermometer with a laser pointer or contact using a K type thermocouple A K type bead...

Страница 5: ...70nm Avoid looking directly at the laser pointer U S law prohibits pointing a laser beam at aircraft doing so is punishable by a fine of up to 10 000 and imprisonment The laser may cause discomfort if...

Страница 6: ...positions of all controls and connectors and the meaning of all display indications and icons before moving on to the Setup Instructions and Operating Instructions 1 Laser pointer 2 Infrared sensor 3...

Страница 7: ...STRUCTIONS INSTALL BATTERY The OS730K s battery compartment Fig 1 Callout 4 is below the measurement trigger and accessible from the front of the grip Before installing the 9V battery included in the...

Страница 8: ...rement is about to be automatically stored recorded in the OS730K s nonvolatile memory IV The measured temperature will also appear on the secondary smaller readout following the brief appearance of a...

Страница 9: ...ckly determine the hottest or coldest object of a group or the difference between their temperatures should be obvious These four display modes MAX MIN AVG and DIF can be entered from REC mode by pres...

Страница 10: ...its emissivity To maximize the accuracy of OS730K measurements you should enter the actual emissivity of the target surface see Appendix II using the instrument s three buttons and the procedure belo...

Страница 11: ...gins flashing you have 30 seconds to adjust its value up or down to any temperature within the unit s measurement range Press the C F button to decrease the value or the button to increase it To move...

Страница 12: ...alues associated with tags 85 and 86 would likely be irrelevant because they represent measurements made nearly 100 hundred trigger squeezes ago USING THE K TYPE PROBE TO MEASURE TEMPERATURE To measur...

Страница 13: ...g the OS730K take the average temperature of all objects within a circular target area Depending on the distance to the target the target area may include both the target and background objects in the...

Страница 14: ...g 1 Callout 2 often but never use a solvent Abrupt temperature changes will cause condensation and possible vapor penetration Clean the LCD after the vapor evaporates Blow off loose particles with cle...

Страница 15: ...the target All IR thermometers IRTs including the OS730K take the average temperature of all objects within a circular target area spot Although the distance D in the D S ratio is defined as a linear...

Страница 16: ...area at the measurement distance the following equation gives the average temperature of the target area Tavg Tmotor Tbackground 2 Solving for Tavg we get 200 75 2 or 137 5 F which is what the OS730K...

Страница 17: ...78 to 0 82 Concrete 0 95 Paint 0 80 to 0 95 Asbestos 0 95 Plastic 0 85 to 0 95 Ceramics 0 95 Paper 0 70 to 0 94 Brass 0 50 Plaster 0 80 to 0 90 Brick 0 90 Rubber 0 95 Carbon 0 85 Wood 0 90 Sludge 0 9...

Страница 18: ...per specification misapplication misuse or other operating conditions outside of OMEGA s control Components in which wear is not warranted include but are not limited to contact points fuses and triac...

Страница 19: ...bine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industri...