UNRESTRICTED
ALARM OPERATION WITH HYSTERESIS & AUTO. RESET (OS-MINI & OS-MINI22)
DATA LOGGING (-MA-R-D, -C4-R-D, -CRT, -BRT MODELS)
The OS-MINI and OS-MINI22 can be used as standalone data loggers. OS-MINI models MA-R-D and C4-
R-D as well as OS-MINI22 models -CRT and -BRT include a MicroSD card slot for data logging, which can
be configured via the touch screen interface. The user can select the sample rate and the number of
samples to be taken and schedule the data logging to start at a certain time.
With a 2 GB card, the user can store 28.4 million readings, which provides almost 1 years’ worth of data at
the fastest possible sample rate of 1 per second.
Data is stored on the MicroSD card in .csv format and can
be viewed and edited easily using spreadsheet software.
A MicroSD card with SD card adapter is available
as an optional accessory.
The MicroSD card slot and battery holder are located on the touch screen circuit board in the lid of the OS-
MI
NI. Readings are time and date stamped using the sensor’s internal clock. The clock is reset when the
power is disconnected, or it will continue if the optional battery is fitted.
DATA LOGGING SPECIFICATIONS
Data Logging Interval
1 to 86,400 seconds (1 day)
MicroSD Card
Max. capacity: 2 GB (not included)
Internal Clock Battery
1 x BR 1225 3V (not included)
Variables Logged
Target temperature, sensing head temperature, electronics module
temperature, max, min, average, emissivity setting, reflected energy
compensation temperature
File Format
.csv
Configurable Parameters
Sample period, number of samples, scheduled start date and time
Modbus address range
1 to 247
USING THE OS-MINI OR OS-MINI22 AS A DATA LOGGER
1.
Insert a MicroSD card into the holder on the circuit board inside the lid of the electronics module.
2.
To retain the date and time when the unit is switched off, fit a battery to the holder on the circuit board
inside the lid.
3.
Replace the lid and connect the sensor power supply.
4.
To set the number of samples to be logged, the time period between samples, and, if required, to
schedule data logging to automatically start, press to access the Settings menu, then press to
access the Data Logging options.
5.
To save data logging settings, press
6.
To manually start data logging, press on the Temperature View or List View.
7.
While logging is in progress, the logging icon flashes on the Temperature View and List View.
8.
To stop data logging, press