background image

OM-CP-TEMP101A

Downloading data from a data logger

- Highlight the data logger in the Connected Devices list. Click “

Stop

” on the menu bar.

- Once the data logger is stopped, with the logger highlighted, click “

Download

”. You will be prompted to 

name your report.

- Downloading will offload and save all the recorded data to the PC. 

Device Maintenance
Battery Replacement

Materials: 
Small Phillips Head Screwdriver 
OM-CP-BAT105 battery

- Puncture the center of the back label with the screw driver and unscrew the enclosure.
- Remove the battery by pulling it perpendicular to the board.
- Insert the new battery into the terminals and verify it is secure.
- Screw the enclosure back together securely.

Recalibration 

Standard calibration is one point at 25 °C.

Additional Services:

 

Custom calibration and verification point options available, please call for pricing.

 
Call for custom calibration options to accommodate specific application needs.

Отзывы: