![Omega iDRX-ACC Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/omega/idrx-acc/idrx-acc_user-manual_3915078031.webp)
Page 27
iDRX-ST SPECIFICATIONS
INPUT RANGE:
±30mV/ ±100mV
Ratiometric or nonratiometric
EXCITATION:
10V @ 30mA or external
INPUT IMPEDANCE:
More than 1MΩ
ISOLATION:
Dielectric strength per 1 minute test based on
EN 61010 for 50 Vdc or Vrms working voltage.
Three way Isolation:
• Power to input 1800V Peak
• Power to analog output/communication 1800V Peak
• Input to analog output/communication 1400V Peak
INPUT OVER-RANGE PROTECTION:
Voltage input: From -10 to +20V
ANALOG TO DIGITAL TECHNIQUE:
Multiple slopes
READ RATE:
8 readings/second, automatic polarity
ACCURACY AT 25 °C:
±0.2% of FS ± 2 counts
TEMPERATURE STABILITY:
75 ppm/°C typical
STEP RESPONSE FOR RS485 OUTPUT:
2 seconds to 99% of the final value
(Filter time constant = 64)
RESPONSE TIME:
To verify the response time, check the carriage return
<CR>, it will be sent at the end of the response. You can
send another command after you receive the <CR>.
i.e.
send: *01X01
response: 01X01<DATA><CR>
Note: 01 is address.
WARM UP TO RATED ACCURACY:
30 minutes
INPUT POWER:
10 to 32 Vdc
POWER CONSUMPTION:
3 watts (125 mA at 24Vdc)
OPERATING AMBIENT:
-5 to +55 °C
STORAGE TEMPERATURE:
-40 to +85 °C
RELATIVE HUMIDITY:
90% at 40 °C non-condensing