Operating Instructions
3-8
a.
Alternatively, to disregard the new offset value and exit the Setup Menu,
press
.
9.
OFFSET is displayed at the top-left of the LCD display.
To set the probe offset when using a calibrated temperature probe with a known offset:
1.
Press
(1.5s)
to enter the Setup Menu;
2.
Press
three (3) times to cycle to the Channel 1 Offset parameter;
3.
Observe the current offset value displayed on the second line of the display;
4.
Press
to set the offset in 0.1 ° increments until the displayed offset value
equals the calibrated probe offset value;
5.
Press
to save the offset value and proceed to Channel 2 offset (HH912T only),
or press
to save the offset value and exit the Setup Menu.
a.
Alternatively, to disregard the new offset value and exit the Setup Menu,
press
.
6.
OFFSET is displayed at the top-left of the LCD display.
Open Wire Detection On/Off
Open Wire Detection allows the unit to detect if a thermocouple probe is connected to the
thermometer. This feature is not compatible with some thermocouple calibrators and can
result in measurement instability.
Turning Off Open Wire Detection in these situations can significantly improve reading
stability. Once off, Open Wire Detection will remain off until changed by following the below
steps, or the instrument is powered off.
If no thermocouple probe is connected and Open Wire Detection is disabled, the unit
will not indicate OPEn and may display erratic readings.
To change the Open Wire Detection setting:
1.
Press
(1.5s)
to enter the Setup Menu;
2.
Press
four (4) times for HH911T, five (5) times for HH912T, to cycle to the
Open Wire Detection Off/On parameter;
3.
“OWD OFF” is flashing near the bottom of the LCD display, and the current Open
Wire Detection status is displayed on Line 2.
4.
Press
to change the Open Wire Detection setting as shown on Line 2 of the
display;
a.
ON indicates that Open Wire Detection is enabled;
b.
OFF indicates that Open Wire Detection is disabled;
5.
Press
or
to save the Open Wire Detection setting and exit the Setup Menu.