25
Enter Backdoor mode by typing:
!11,MW,1000,1[CR]
Unit will respond with:
!11,BackDoorEnabled: Y
Disable DAC update by typing:
!11,WRITE,4,D[CR]
Unit will respond with:
!11,DisableUpdate: D
7.3.2 Gas flow 0-5 Vdc analog output calibration
1.
Install jumpers J7A, J7B and J7C on the PC board for 0-5 Vdc output (see Table VI).
2.
Connect a certified high sensitivity multi meter set for the voltage measurement to the
pins 2 (+) and 1 (-) of the 15 pins D connector.
3.
Write 4000 counts to the DAC channel 1: !11,WRITE,1,4000[CR]
4.
Read voltage with the meter and calculate:
5.
Save FlowOutScaleV in to the EEPROM:
!11,MW,25,X[CR]
Where: X – the calculated AoutScaleV value.
7.3.3 Gas flow 4-20 mA analog output calibration
1. Install jumpers J7A, J7B and J7C on the PC board for 4-20 mA output (see Table VI).
2.
Connect a certified high sensitivity multi meter set for the current measurement to
pins 2 (+) and 1 (-) of the 15 pins D connector.
3.
Write 4000 counts to the DAC channel 1:
!11,WRITE,1,4000[CR]
4.
Read current with the meter and calculate:
5.
Write zero counts to the DAC channel 1:
!11,WRITE,1,0CR]
6.
Read offset current with the meter and calculate:
7. Save AoutScale_mA in to the EEPROM:
!11,MW,27,Y[CR]
Save AoutOffset_mA in to the EEPROM:
!11,MW,28,Z[CR]
Where: Y – the calculated AoutScale_mA value.
Z – the calculated AoutOffset_mA value.
Note:
When done with the analog output calibration make sure the
DAC update is enabled and the BackDoor is closed
(see command below).
Содержание FMA 4000
Страница 1: ...www omega com e mail info omega com User sGuide FMA 4000 Digital Mass Flow Meters Shop online at...
Страница 50: ...46 APPENDIX IV COMPONENT DIAGRAM TOP COMPONENT SIDE Aug 09 2007...
Страница 51: ...47 BOTTOM COMPONENT SIDE Aug 09 2007...
Страница 52: ...APPENDIX V DIMENSIONAL DRAWINGS FMA 4000 WITHOUT READOUT 48...
Страница 53: ...FMA 4000 WITH READOUT OPTION 49...