6-14
1
2
3
4
1
2
3
4
<EOI>
<EOI>
For Measured Data
For Computed Data
Measured data 1 (first CH)
Measured data 1 (last CH)
Measured data 2 (first CH)
Measured data 2 (last CH)
Measured data n (first CH)
Measured data n (last CH)
Computed data 1 (first CH)
Computed data 1 (last CH)
Computed data 2 (first CH)
Computed data 2 (last CH)
Computed data n (first CH)
Computed data n (last CH)
Number of Output Bytes
Number of Output Bytes
The above number of output bytes is indicated as follows:
Measured data:
2 x channel number x n byte (n: number of data items per channel)
Computed data:
4 x channel number x n byte (n: number of data items per channel)
Invalid data are output as follows:
Plus over data
7FFFH (7FFF7FFFH for computed data)
Minus over data
8001H (80018001H for computed data)
Measurement range setting is SKIP
8002H (80028002H for computed data)
Abnormal data
8004H (80048004H for computed data)
No data
8005H (80058005H for computed data)
Note
• The output data are all in hexadecimal format.
• Measured data can be output with either the most significant byte or least significant byte following the order of output
bytes with the BO command.
• If the RAM disk is faulty, the character string “# ERROR ON DATA MEMORY” will be output.
• If there are no appropriate files, the character string “# FILE NAME ERROR” will be output.
6.11 RAM Disk Output Format (Binary Code)
Содержание DR130 SERIES
Страница 1: ...http www omega com e mail info omega com User sGuide DR130 DR230 DR240 SERIES Communication Interface ...
Страница 39: ...4 9 Command Format 4 4 5 Input Range Parameter This page intentionally left blank ...
Страница 40: ...4 10 4 5 Input Range Parameter This page intentionally left blank ...
Страница 56: ...5 15 Commands 5 This page intentionally left blank 5 5 Other Settings ...
Страница 71: ...6 9 Output Format 6 This page intentionally left blank ...