4
Configuring The Meter
28
4.3.2 Scaling Without Known Loads (Continued)
3.
Determine
IN1
&
IN!2
input range and resolution. The example selects the 0 to
10 V range and 10 uV resolution (
R.2=4
).
Example:
IN1
= (1 Volt) x (1000 cts/v) x (1.000) = 1000
IN!2
= (5 Volt) x (1000 cts/v) x (1.000) = 5000
RD1
= 0000
RD!2
= 2000
4.
Press
MENU
button until the meter shows
RD.S.O
.
5.
Press the
∂
∂
TARE
button. The meter shows
IN1
.
6.
Press the
∂
∂
TARE
button again, the meter shows the last Input 1 value, with the
fourth digit flashing.
7.
Press the
ß
ß
MAX
button to change the value of your digits.
8.
Press the
∂
∂
TARE
button to scroll horizontally to the next digit.
9.
Press the
MENU
button to store this value. The meter shows
RD1
.
10.
Press the
∂
TARE
button. The meter shows the last value for read 1.
Repeat steps 7, 8 and 9 until
RD1
,
IN!2
and
RD!2
have been displayed, verified,
changed (if necessary) and stored.
4.4 USING READING CONFIGURATION
RD.CF
Refer to Table 6-1 for a summary list of menu configuration.
You may use Reading Configuration
RD.CF
to configure your meter for the following:
•
To enable or disable the front panel tare
•
To set the input resolution of your meter
•
To display the filtered/unfiltered signal input value
Note
☞
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание DP25B-E
Страница 11: ...viii Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 70: ...59 NOTES Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 71: ...60 NOTES Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...