9) After filling the electrolyte solution and screw the membrane cap, please find the insulation tape to seal the

seam of the sensor head to reduce solution leakage, please refer to the picture below:

Insulation tape 

10) Check the membrane module to see whether there are air bubbles in it. If found there are air bubbles, knock on the

membrane cap carefully to dispel them.

11) Inspect the membrane to ensure that the inner cathode element contacts the membrane. The membrane must be

taut, without wrinkles or imperfections. <If wrinkles are found, please follow 

2) 

to change the membrane.

12) Connect DO electrode with Meter DO BNC connector.
13) Rinse the assembled electrode with distilled water and blot dry with a clean lab wipe.

C) DO calibration C-1) or C-2) or C-3) :

There are three ways to do the DO calibration, for common and convenient way, follow c-1. 
For more accurate measurement, follow c-2 and c-3 with necessary devices and solution in laboratories. 

NOTE: 

First doing zero DO calibration and then doing 100% air-saturated water calibration. 

C-1) 100% Water-Saturated Air Calibration: 

(Convenient for self-calibration, common used)

1) Connect the electrode to the meter.
2) By positioning the sensor in air-saturated water (air is directed through water until the water is saturated

with it). The illustration below is a representation of the conditions in air- saturated water.

3) Press and hold 

MODE+CAL 

buttons to enter calibration mode, screen shows “DO %100” then press and hold

SET 

button to save and screen shows “

SA

” to complete calibration. Press and hold 

MODE+CAL 

buttons, screen

shows “

ESC

” momentarily and return to normal measurement mode.

Содержание DOH-10 Series

Страница 1: ...ser sGuide Shop online at omega com e mail info omega com For latest product manuals omega com en us pdf manuals DOH 10 Series Handheld Dissolved Oxygen Meter Kit with Optional SD Card Data Logger TM...

Страница 2: ...78 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide The information contained in...

Страница 3: ...l testing lakes streams oceans Water Wastewater treatment Wine production Tripod receptacle mountable design for long time monitoring purposes 1 Meter 2 Battery AAA x 3 pcs 3 Electrode x 1pcs DO Galva...

Страница 4: ...g mode UNIT Short press to change temperature unit Long press to enter temperature electrode type selection Short press to select NTC Negative Temperature Coefficient NOT no remote temperature electro...

Страница 5: ...rwise LCD shows erratic readings and leakage may occur Warranty is void if not follow the notices Note Remove batteries when not use Power off switch When the meter is turned off the internal CPU does...

Страница 6: ...used B DO electrode installation NOTE DO NOT touch the sensitive membrane when using the electrode or replacing the membrane cap for the sweat and grease will affect the quality of membrane and decre...

Страница 7: ...Push the membrane cap downward gently until it won t lose 7 Fill the membrane cap with the DO electrolyte solution 0 5M NaOH provided Some solution will be vented from the overflow port This is norma...

Страница 8: ...n lab wipe C DO calibration C 1 or C 2 or C 3 There are three ways to do the DO calibration for common and convenient way follow c 1 For more accurate measurement follow c 2 and c 3 with necessary dev...

Страница 9: ...r and wait for the reading to stabilize press and hold MODE CAL buttons to enter calibration mode screen shows DO 100 then press and hold SET button to save and screen shows SA to complete calibration...

Страница 10: ...ce is constantly replenished A moving stream will provide adequate circulation Freeze the current readings of DO and temperature readings by press HOLD button then icon Hold will appear on the top lef...

Страница 11: ...2 5 08 15 0 10 08 9 54 9 03 8 54 8 08 7 64 41 0 6 31 6 03 5 75 5 49 5 24 5 01 16 0 9 87 9 34 8 84 8 37 7 92 7 5 42 0 6 21 5 93 5 67 5 41 5 17 4 93 17 0 9 67 9 15 8 67 8 21 7 77 7 36 43 0 6 12 5 84 5 5...

Страница 12: ...ust connected to your computer Right click on your SD card to bring up the right click menu options Select Format Keep Capacity and Allocation unit size set to be default c Select the file system This...

Страница 13: ...serted 4 When an SD card is used for the first time a folder is created on the card and named with the model number Under the MODEL number folder the MODEL number and AUTO YEAR folder will be automati...

Страница 14: ...ou want to keep the previous data rename the file MANUAL csv in DOH 10 MANUAL csv is required 4 Data directory in SD card DOH 10 MANUAL csv NOTE When manual data records full 199 points logging will c...

Страница 15: ...rodes When not in use Long term For long term storage or to take the electrode out of service disconnect the electrode from the meter Disassemble the electrode membrane cap Rinse the anode cathode and...

Страница 16: ...se shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OME...

Отзывы: