![Omega CNi16-AL Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/omega/cni16-al/cni16-al_user-manual_736770012.webp)
2.3.2 Thermocouple
The figure below shows the wiring hookup for any thermocouple type. For
example, for Type K hookup, connect the yellow wire to the "2" terminal and the
red wire to the "1(-)" terminal.
When configuring your monitor, select Thermocouple and Thermocouple
Type in the Input Type menu (see Part 3).
Figure 2.5 Thermocouple Wiring Hookup
8
TYPE
Input Connector
Jacket (external insulation)
Terminal 1 (-)
Terminal 2 (+)
Extension
Grade
J
Red
White
dark-Brown
Black
K
Red
Yellow
dark-Brown
Yellow
T
Red
Blue
dark-Brown
Blue
E
Red
Purple
dark-Brown
Purple
N
Red
Orange
dark-Brown
Brown
R
Red
Black
-
Green
S
Red
Black
-
Green
B
Red
Gray
-
Black
Table 2.4 TC Wire Color Chart