9 COMMUNICATION 

9.1 Communication 

Protocol 

CL551

 is equipped with a RS232 bi-directional , half-duplex interface as standard. 

Baud rate is set at 9600 baud as default. You can modify it through keyboard selecting following rate : 300, 600, 1200, 
2400, 4800, 9600 and 19200 baud. Transmission parameters are 8 bit length, no parity and 1 bit stop. Address can be 
modified through keyboard and it is set 1 as default. RS232 interface connector is a 9 pin DB9 male connector. 
 
Cable connection to a standard personal computer as follows : 
 

9 pin CL551            9 pin PC connector 

25 pin PC connector

 

 
2 ----------------------------------------> 3 --------------------------> 2 

 

 

3 ----------------------------------------> 2 --------------------------> 3 

 

 

5 ----------------------------------------> 5 --------------------------> 7 

 

 

 
Communication protocol is based on Answer to Request system. 

CL551

 will answer only if you sent a Request to it. 

 
There are two kind of request : 

RVAR  to read a variable. 
WVAR to set a variable. 

9.2 

Reading a variable 

The instruction to read a variable has to be an ASCII code composed as follows : 
 
$ character 
Instrument Address 
RVAR request string 
Variable Index 
Blank space 
Carriage Return 
 
For example, if the set point has to be read (Variable index = 0), supposing the instrument address to be 1: 
 
$1RVAR0 \r 
 
ASCII 

code 

$ 1 R V A R 0   \r 

HEX 

code 

24 31 52 56 41 52 30 20 0D 

 
The answer of the 

CL551

 will be as follows  

* character 
Carriage return 
* character 
Instrument Address 
Answer string 
Carriage Return 
 
 
 
For example if set point is 50.5 °C , supposing the instrument address to be 1: 
 

ASCII 

code 

* \r 

*  1   S P    

:   0 0  5 0 .  5 \r 

HEX $ 
code 

2A 0D 

2A 31 20  53 50  20 20  3A 20 30  30  35 30 2E 33 0D 

 
List of Variable Index available 
 

Variable Index 
Actual Setpoint 

Ramp : ON /OFF 

22 

Содержание CL551

Страница 1: ...http www omega com e mail info omega com CL551 Portable Temperature Calibrator...

Страница 2: ...0800 1 66342 e mail czech omega com FAX 420 69 6311114 France 9 rue Denis Papin 78190 Trappes Tel 33 130 621 400 Toll Free in France 0800 4 06342 e mail france omega com FAX 33 130 699 120 Germany Aus...

Страница 3: ...n order to pursue the technological leadership these improvements could require changes to the information of this book OMEGA reserves the right to change such information without notice OMEGA makes n...

Страница 4: ...ice 10 5 4 Thermocouple wires 11 6 SAFETY RECOMMENDATIONS 12 7 OPERATLON APPLICATION 13 7 1 Calibration by comparison 13 7 1 1 Calibration using the internal Pt100 sensor 13 7 1 2 Calibration using an...

Страница 5: ...s CL551 is a transportable unit designed to obtain a controlled temperature with high stability high uniformity in a range from ambient temperature to 550 C The internal metal block designed with a se...

Страница 6: ...jacket Internal metal block standard aluminum with n 6 holes 3 175 4 75 6 5 8 0 9 5 19 4 7 mm Electronic section in a separate metal shell Internal temperature sensor Pt100 Class A sensor control and...

Страница 7: ...or the calibration of the sensor 2 3 Internal digital indicator The internal digital indicator is a dot matrix LCD display with 2 lines of 16 characters equipped with a backlight device for easy readi...

Страница 8: ...e sensor and a modulated air blower The module on the right is endowed with an alphanumeric dot matrix LCD display with a backlight device a microcontroller mother board with all base functions an ope...

Страница 9: ...Standard Sensor POWER SUPPLY Mains supply 110 or 220V MICROCONTROLLER SWITCH TEST KEYBOARD RS 232 INTERFACE Internal reference sensor Equalizing Block Peltier cells The instrument functional blocks a...

Страница 10: ...ds and to ensure good general performances The signal leads should not be run near ac line wiring transformer and heating elements Connection leads should if possible be twisted and shielded with the...

Страница 11: ...f error exceeding those in the table Errors up to 4 C may be experienced when using the Extension grade thermocouple wire for J and K thermocouples Limit of Error of thermocouples The range indicated...

Страница 12: ...hot handle the oven to avoid thermal hazard and make sure that it has cooled down before putting it into its case Do not fill the thermal equalizer block holes with any type of fluid Do not insert the...

Страница 13: ...Y THE PROBE S FROM THE THERMAL EQUALIZING BLOCK WAIT FIRST FOR THE TEMPERATURE CALIBRATOR TO COOL AT THE END OF THE OPERATION WAIT FOR THE THERMAL EQUALIZING BLOCK TO BE COMPLETELY COOLED BEFORE PLACI...

Страница 14: ...instrument Verify the reading temperature of the reference sample with the values reported on the Certificate of Conformity make necessary relative variations Calibrate the value read on the sensor un...

Страница 15: ...rator validates it by pressing the E key An acoustic beep together with a temporary indication right on the screen will indicate that the value has correctly been stored into memory 7 2 1 Display reso...

Страница 16: ...ust the SP to 95 C using the key Press the E key to confirm the value Within a minute the heater temperature will settle to 95 C When the temperature is settled press the F key The display will show T...

Страница 17: ...the following indication relevant to the Proportional Band constant P B 1 6 T 23 5 C The PID constants are experimentally defined and memory loaded for tuning optimization The operator can eventually...

Страница 18: ...ameters security or access code baud rate machine code serial number The 2nd level of the set up procedure starts from the following indication that you can meet at the end of the 1st level of the set...

Страница 19: ...max temperature limit MAX set 550 0 T 23 5 C Press simultaneously F keys to exit this procedure and to return to the normal operative mode with the following indication SP 150 0 T 23 5 C 7 5 P I D tu...

Страница 20: ...c Replace the triac stop at the set point value 4 The set point value isn t the The internal Pt100 or the Check and replace the engaged value or the Microcontroller is damaged part temperature shown o...

Страница 21: ...Tweezers 16 Pt100 external connector 17 RS232 connector 18 Transport handle 19 Internal carpentry 20 Rear carpentry 21 Tuning container 8 3 Storage Store the instrument in the original package at a t...

Страница 22: ...wer only if you sent a Request to it There are two kind of request RVAR to read a variable WVAR to set a variable 9 2 Reading a variable The instruction to read a variable has to be an ASCII code comp...

Страница 23: ...s to be set Variable index 0 to 123 C supposing the instrument address to be 1 1WVAR0 123 r ASCII code 1 W V A R 0 1 2 3 r HEX code 24 31 57 56 41 52 30 20 31 32 33 0D The answer of the CL551 will be...

Страница 24: ...Resolution 0 01 0 1 4 1 for 0 01 0 for 0 1 Proportional band 5 decimal value Integral time 6 decimal value Derivative time 7 decimal value Pt100 Internal External 8 1 for Internal 0 for External Engi...

Страница 25: ...or otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment s...

Страница 26: ...addlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Equipme...

Отзывы: