background image

7  

OPERATlON & APPLICATION 

7.1 

Calibration by comparison 

To verify or calibrate a temperature sensor by  comparison with an internal or  external Pt100 standard  follow the 
procedure indicated below. For correct operations strictly follow the following recommendations: 
 

• 

Measure the outside diameter of the sensor to be tested. 

• 

Insert the sensor to be tested in one hole of the thermal equalizing block. For a correct calibration the outside 
diameter of the sensor under test should be approximately 0.5 mm less than the inside diameter of the hole; 

• 

If required use the appropriate well adapter (see the figure below). 

• 

Do not force the sensor to be tested into the thermal equalizing block (see the figure below).   

 

NO

OK

NO

OK

>0.5 mm

 

 
 

ATTENTION 

 

D

O NOT INSERT THE SENSOR UNDER TEST IN THE THERMAL EQUALIZING BLOCK WHEN THE OVEN IS AT A HIGH TEMPERATURE

T

HERMAL SHOCK COULD DAMAGE THE SENSING ELEMENT OF THE PROBE

A

T THE END OF THE TEST DO NOT REMOVE IMMEDIATELY THE PROBE

/

S FROM THE THERMAL EQUALIZING BLOCK

W

AIT FIRST FOR THE TEMPERATURE CALIBRATOR TO COOL

.  

A

T THE END OF THE OPERATION

 :  

WAIT FOR THE THERMAL EQUALIZING BLOCK TO BE COMPLETELY  COOLED BEFORE PLACING THE 

CALIBRATOR INTO THE CARRYING CASE

7.1.1 

Calibration using the internal Pt100 sensor 

• 

Start the calibration only at ambient temperature. 

• 

Insert the sensor under test in the correct hole with the sensing element into the recommended  calibration zone as 
indicated in the following figure. 

• 

Switch the instrument on using the main switch on the rear panel. 

• 

Pressing the 

<

>

 key the set-point values will increase. 

• 

Pressing the 

<

>

 key the set-point values will decrease. 

• 

To confirm values, press the 

<E>

 key until a sound is made by unit. 

• 

Verify the reading temperature on the display with the values reported on the Certificate of Conformity and make the 
relative modifications. 

• 

Calibrate the value read on the sensor under test with the correct value on the indicator. The check can be 
accomplished with several temperature values. In order to follow this procedure just set up the set-point to the 
relevant temperature value and wait for stabilization. 

 

13 

Содержание CL551

Страница 1: ...http www omega com e mail info omega com CL551 Portable Temperature Calibrator...

Страница 2: ...0800 1 66342 e mail czech omega com FAX 420 69 6311114 France 9 rue Denis Papin 78190 Trappes Tel 33 130 621 400 Toll Free in France 0800 4 06342 e mail france omega com FAX 33 130 699 120 Germany Aus...

Страница 3: ...n order to pursue the technological leadership these improvements could require changes to the information of this book OMEGA reserves the right to change such information without notice OMEGA makes n...

Страница 4: ...ice 10 5 4 Thermocouple wires 11 6 SAFETY RECOMMENDATIONS 12 7 OPERATLON APPLICATION 13 7 1 Calibration by comparison 13 7 1 1 Calibration using the internal Pt100 sensor 13 7 1 2 Calibration using an...

Страница 5: ...s CL551 is a transportable unit designed to obtain a controlled temperature with high stability high uniformity in a range from ambient temperature to 550 C The internal metal block designed with a se...

Страница 6: ...jacket Internal metal block standard aluminum with n 6 holes 3 175 4 75 6 5 8 0 9 5 19 4 7 mm Electronic section in a separate metal shell Internal temperature sensor Pt100 Class A sensor control and...

Страница 7: ...or the calibration of the sensor 2 3 Internal digital indicator The internal digital indicator is a dot matrix LCD display with 2 lines of 16 characters equipped with a backlight device for easy readi...

Страница 8: ...e sensor and a modulated air blower The module on the right is endowed with an alphanumeric dot matrix LCD display with a backlight device a microcontroller mother board with all base functions an ope...

Страница 9: ...Standard Sensor POWER SUPPLY Mains supply 110 or 220V MICROCONTROLLER SWITCH TEST KEYBOARD RS 232 INTERFACE Internal reference sensor Equalizing Block Peltier cells The instrument functional blocks a...

Страница 10: ...ds and to ensure good general performances The signal leads should not be run near ac line wiring transformer and heating elements Connection leads should if possible be twisted and shielded with the...

Страница 11: ...f error exceeding those in the table Errors up to 4 C may be experienced when using the Extension grade thermocouple wire for J and K thermocouples Limit of Error of thermocouples The range indicated...

Страница 12: ...hot handle the oven to avoid thermal hazard and make sure that it has cooled down before putting it into its case Do not fill the thermal equalizer block holes with any type of fluid Do not insert the...

Страница 13: ...Y THE PROBE S FROM THE THERMAL EQUALIZING BLOCK WAIT FIRST FOR THE TEMPERATURE CALIBRATOR TO COOL AT THE END OF THE OPERATION WAIT FOR THE THERMAL EQUALIZING BLOCK TO BE COMPLETELY COOLED BEFORE PLACI...

Страница 14: ...instrument Verify the reading temperature of the reference sample with the values reported on the Certificate of Conformity make necessary relative variations Calibrate the value read on the sensor un...

Страница 15: ...rator validates it by pressing the E key An acoustic beep together with a temporary indication right on the screen will indicate that the value has correctly been stored into memory 7 2 1 Display reso...

Страница 16: ...ust the SP to 95 C using the key Press the E key to confirm the value Within a minute the heater temperature will settle to 95 C When the temperature is settled press the F key The display will show T...

Страница 17: ...the following indication relevant to the Proportional Band constant P B 1 6 T 23 5 C The PID constants are experimentally defined and memory loaded for tuning optimization The operator can eventually...

Страница 18: ...ameters security or access code baud rate machine code serial number The 2nd level of the set up procedure starts from the following indication that you can meet at the end of the 1st level of the set...

Страница 19: ...max temperature limit MAX set 550 0 T 23 5 C Press simultaneously F keys to exit this procedure and to return to the normal operative mode with the following indication SP 150 0 T 23 5 C 7 5 P I D tu...

Страница 20: ...c Replace the triac stop at the set point value 4 The set point value isn t the The internal Pt100 or the Check and replace the engaged value or the Microcontroller is damaged part temperature shown o...

Страница 21: ...Tweezers 16 Pt100 external connector 17 RS232 connector 18 Transport handle 19 Internal carpentry 20 Rear carpentry 21 Tuning container 8 3 Storage Store the instrument in the original package at a t...

Страница 22: ...wer only if you sent a Request to it There are two kind of request RVAR to read a variable WVAR to set a variable 9 2 Reading a variable The instruction to read a variable has to be an ASCII code comp...

Страница 23: ...s to be set Variable index 0 to 123 C supposing the instrument address to be 1 1WVAR0 123 r ASCII code 1 W V A R 0 1 2 3 r HEX code 24 31 57 56 41 52 30 20 31 32 33 0D The answer of the CL551 will be...

Страница 24: ...Resolution 0 01 0 1 4 1 for 0 01 0 for 0 1 Proportional band 5 decimal value Integral time 6 decimal value Derivative time 7 decimal value Pt100 Internal External 8 1 for Internal 0 for External Engi...

Страница 25: ...or otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment s...

Страница 26: ...addlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Equipme...

Отзывы: