17
calibration value. This function is generally used for replace with new
sensor or data confusion. Generally after restored factory setting, it need
recalibration before using.
5. ERROR CODES
When the instrument detects an error condition, an error code will be
displayed. All the error codes are described below.
6. SENSOR
We use foreign advance technology to manufacture our conductivity sensor.
The quality of the sensor is excellence and it can be used in industry
province with all kind of conductivity monitor.
6.1 Sensor Features
High sensitivity and accuracy with reliability and durability
Stainless material
Quickly response (within two seconds) and stability
Unique structure preventing the sensor from pollution and blockage
Easy to assemble and clean
Code
Content
Solve methods
Er02
Measurement
value out of range
Check whether the
choice of cell constant
is suitable and replace the correct one
Er03
Temperature
sensor failure
1. Check whether the temperature sensor
connected with the monitor properly
2. Check whether the temp. compensation
(menu CC) setting correct
Table 9 Error codes