15
Table 6
Set current output
4.3.3.1 FSIS
、
FSIE—set current output start and end value
Select FSIS in the menu, press
ENTER
into it and display the original
current output start value, use
DOWN
and
UP
to modify it, then press
ENTER
to store and return to the setting state.
The same method select FSIE
to set the current output end value.
4.3.3.2 FSI—select the input for the current output
Select FSI in the menu, press
ENTER
into it and display FS 0 or FS 1. FS 0
is conductivity iutput, FS 1 is temperature as the input. Press
UP
to
switch it, then Press
ENTER
to store and return to the setting state.
Code
Content
Direction
FSIS
Current output start value
4.00~20.00 mA
FSIE
Current output end value
4.00~20.00 mA
FSI
The input for the current
output
0: conductivity
1: temperature
4.3.3.3 FIgS
、
FIgE
、
FICS
、
FICE—set input value range for the current
output
After selecting the input parameters, you can set its start value and end
value. Select FIgS in the menu, press
ENTER
into it and display the original
conductivity start value, use
DOWN
and
UP
to modify it, then press
ENTER
to store and return to the setting state. The same method select
FIgE
to set the conductivity end value.
Likewise, select FICS and FICE to set the temperature start and end value for
current output.