7
SY
Sets or queries whether the mode is normal mode or calibration
mode
Normal Cal.
Command = SYm<CRLF> -> Answer = SYm<CRLF>
Command = SY?<CRLF> -> Return = SYm<CRLF>
Parameters
m = 0: Normal mode (default)
1: Calibration mode
UP
Increases the m
th
digit of the output value by 1
Normal Cal.
Command = UPm<CRLF> -> Answer = UP,OK<CRLF>
Parameters
m = 1 to 5 (1 is the least significant digit while 5 is the most
significant digit)
If the CA450 is in any mode other than constant current output
mode, ERR13 is returned.
If the CA450 is in span check mode, ERR13 is returned.
Î
UQ
Increases the output value by one step or sets the output value
to 100%
Normal Cal.
Command = UQ<CRLF> -> Answer = UQ,OK<CRLF>
Increases the output value by one step (25%).
Sets the output value to 100% of the span in span check mode.
If the CA450 is in any mode other than constant current output
mode, ERR13 is returned.
Î