PRTXB Series RTD Process Thermometers

3

Calibration

The PRTXB is factory calibrated and there is generally no need 
to alter calibration settings. Required calibration equipment 
includes a temperature reference of at least four times the 
unit’s accuracy, a dry-block calibrator or a temperature 
controlled bath. 
Calibration may be performed in any of the available 
temperature units, the use of Fahrenheit or Celsius is assumed 
in this procedure. Select the temperature units for calibration 
prior to entering the calibration mode. 
The unit enters and remains in the calibration mode until 
restarted manually or power is removed. While in the calibration 
mode, the auto shutoff timer is disabled, and the Min/Max 
feature is disabled.
The unit is calibrated at two points, at ice point and at a 
temperature above ice point. 
For general service, the full scale temperature is normally used 
for the second point. However, if a particular temperature is of 
critical interest it may be used instead for greatest accuracy at 
that point. 

Enter Calibration Mode 

To enter the calibration mode, begin with 
the unit powered off, and press and hold the 

t

 button.

Then press the power button.
Release all buttons when the display 
indicates CAL.
Before the unit enters the calibration mode, 
the display initially indicates _ _ _ _ with the 
first underscore blinking, and with CALPC on 
the lower display. 
Note: The unit will automatically revert to 
normal operation if no buttons are operated 
for approximately 15 seconds.
To cancel and return to normal operation, 
press and release the power button without 
entering any pass code characters.
Enter the user-modifiable pass code (3510 
factory default).
Use the 

s

 and 

t

 buttons to set the left-

most digit to 3.
Press and release the power button to index 
to the next position. The 3 will remain, and 
the second position will be blinking.
Use the 

s

 and 

t

 buttons to select 5.

Press and release the power button to index 
to the next position. 3 5 will remain, and the 
third position will be blinking.
Use the 

s

 and 

t

 buttons to select 1.

Press and release the power button to index 
to the next position. 3 5 1 will remain, and 
the fourth position will be blinking.
Use the 

s

 and 

t

 buttons to select 0.

Press and release the power button to 
proceed with calibration.
Note: If an incorrect pass code is entered, the 
unit will return to the start of the pass code 
entry sequence.

Calibration Procedure

Upon successful pass code entry, the 
upper display will indicate the RTD probe 
temperature. The lower display will alternate 
as indicated below.
Note: To store the calibration parameters 
and exit calibration mode at any time, press 
and hold the power button until the display 
indicates  - - - -.

Ice-Point Calibration

When the applied temperature is below 
approximately 12 °C (or 54 °F), the unit will 
automatically select the ice-point calibration 
mode.
Apply 0.0 °C or 32.0 °F to the RTD.
The lower display will alternate between ICE 
and DEG C or DEG F. 
Use the 

s

 and 

t

 buttons to adjust the upper 

display to indicate 0.0 °C or 32.0 °F.

Span Calibration

Apply full-scale temperature to the RTD. 
The lower display segments will alternate 
between CAL and DEG C or DEG F. 
Use the 

s

 and 

t

 buttons to adjust the upper 

display segments to indicate the applied 
temperature value.
To store the calibration parameters and exit 
calibration mode, press and hold the power 
button until the display indicates  - - - - .

Changing the Pass Codes

The factory default pass code of 3510 may be changed if 
desired. Separate pass codes may be used for min/max 
configuration access and calibration access. 

Configuration Pass Code Access

With the unit off, press and hold the 

s

 

button.
Then press the power button.
Release all buttons when the display 
indicates CFG.

Calibration Pass Code Access

With the unit off, press and hold the 

t

 button 

to view and/or change the user calibration 
pass code.
Then press the power button.
Release all buttons when the display 
indicates CAL.

View/Change Pass Code

Before the unit enters the view or change 
pass code mode, the display initially indicates  
_ _ _ _  with the first underscore blinking, 
and with CFGPC or CALPC on the lower 
display. 
Note: The unit will automatically revert to 
normal operation if no buttons are operated 
for approximately 15 seconds.
To cancel and return to normal operation, 
press and release the power button without 
entering any pass code characters.

Enter Access Code 1220

Use the 

s

 and 

t

 buttons to set the left-

most digit to 1.
Press and release the power button to index 
to the next position. The 1 will remain, and 
the second position will be blinking.
Use the 

s

 and 

t

 buttons to select 2.

Press and release the power button to index 
to the next position. 1 2 will remain, and the 
third position will be blinking.
Use the 

s

 and 

t

 buttons to select 2.

Press and release the power button to index 
to the next position. 1 2 2 will remain, and 
the fourth position will be blinking.
Use the 

s

 and 

t

 buttons to select 0.

Note: If an incorrect access code was 
entered, the unit will return to the start of the 
access code entry sequence.

Press and release the power button to 
proceed.
The display will indicate the existing user-
defined pass code with CFGPC or CALPC on 
the lower display. To exit without changes, 
press the power button.
Operate the 

s

 and 

t

 button to select 

the first character of the new pass code. 
Characters 0-9 and A, b, C, d, E, F may be 
used.
When the correct first character is being 
displayed, press and release the power 
button to proceed to the next pass code 
character.
Repeat above until the entire pass code is 
complete.
To exit the view or change pass code mode, 
press and hold the power button. 
Release the button when the display 
indicates - - - - to restart the unit.
 

- - - -

t

s

*

Press to save and 

restart

 ice

32.0

t

s

 deg F

*

*

or

Increment up 

or down

 cal

392.0

t

s

*

*

or

Increment up 

or down

 deg F

- or -

cf6

t

s

Release buttons

t

s

*

2

*

1

cfgpc

____

t

s

cfgpc

i___

t

s

*

*

or

*

Increment 

up or down

Move to 

next #

cfgpc

i220

t

s

*

*

or

*

Increment 

up or down

Move to 

next #

cal

t

s

Release buttons

t

s

*

2

calpc

____

t

s

*

1

cal

t

s

Release buttons

t

s

*

2

*

1

calpc

____

t

s

calpc

3___

t

s

*

*

or

*

Increment 

up or down

Move to 

next #

calpc

35 i0

t

s

*

*

or

*

Increment 

up or down

Move to 

next #

calpc

35 i0

t

s

*

Press to enter 

calibration mode

def F

75.0

t

s

 cal

cfgpc

35 i0

t

s

*

*

or

*

Increment 

up or down

Move to 

next #

cfgpc

A250

t

s

*

*

*

Increment 

up or down

Move to 

next #

cfgpc

A250

t

s

*

Press to save and 

restart

cfgpc

35 i0

t

s

*

Press to enter pass 

code configuration 

mode