The PRTXB series is microprocessor controlled industrial RTD 
temperature indicator with a digital temperature display in a 
rugged NEMA 4X housing.
The temperature reading is linearized for the digital display. The 
temperature display may be set up to read °F, °C, or kelvin and 
the auto shutoff time may be set as needed.
The unit is capable of automatically capturing and storing 
maximum and maximum readings. The min/max functionality 
can be set up as required, and the readings can be either saved 
or cleared when the unit shuts off.

Installation and Precautions

Read these instructions before installation. Configuration may 
be easier before installation. 
The spring-loaded versions must be used with a thermowell. 
Use a thermowell appropriate for the process. 
The non-spring-loaded versions can be used in non-pressurized 
applications or applications with no flow. Due to the hardness of 
316 stainless steel, it is recommended that a thread sealant be 
used to ensure leak-free operation. 
Do not exceed maximum allowable housing temperature. 
Install or remove using wrench on probe hex fitting only. Do not 
attempt to tighten or loosen by turning the housing.

Operation

Press and release the power button to power 
up the unit. The unit tests all LCD segments 
and displays the RTD temperature on the 
upper display and the temperature units on 
the lower display. Readings are updated 
approximately 4 times per second.
The RTD probe has a time constant of 
approximately 10 seconds, typical of an RTD 
probe in a stainless sheath. Time constant is 
characterized as the RTD changing to 63.2% 
of its new temperature span in one time 
constant, and 95% of its new temperature 
span in three time constants.
If the unit is configured with an auto shutoff 
time, a five second warning period is 
provided prior to auto shutoff, during which 
the display indicates OFF. The auto shutoff 
timer is reset whenever any button is pressed 
and released. 
To shut off the unit manually at any time, 
press and hold the power button until the 
display indicates OFF (up to about 5 seconds 
total if MAX/MIN is enabled) and then release 
the button.

Out-of-Range Indications

If the RTD temperature goes above 392°F or 
200°C, ALARM1 will be displayed. 
If the RTD temperature goes below –58°F or 
–50°C, ALARM2 will be displayed. 
If the RTD temperature continues beyond 
these limits, the display will eventually 
indicate 1.-.-.-.

MAX/MIN Operation (if enabled)

The factory default setting has min/max disabled. To turn this 
feature on, see the section titled User Configuration Mode. The 
unit may be configure to use max, min, neither, or both.
To step the unit through the display modes, 
press and hold the power button about 1 
second until the display indicates MAX or 
MIN and then release the button. The display 
mode cycle repeats through the following 
steps.

MAX Mode 

The display indicates the stored maximum 
reading and the lower display alternates 
between indicating MAX and the temperature 
units. The thermometer may be left in this 
mode if desired.

MIN Mode

Press and hold the power button about 1 
second The display indicates the stored 
minimum reading and the lower display 
alternates between indicating MIN and the 
temperature units. The thermometer may be 
left in this mode if desired.

Return to Normal Mode

To return from minimum or maximum 
reading mode to the normal operating 
mode, press and hold the power button for 
about 1 second until the display indicates 
the temperature units and then release the 
button. The unit returns to the normal mode. 
The stored maximum and minimum values 
are not cleared and continue to be updated. 

Reset MAX and MIN

To manually reset the stored maximum 
and minimum readings, press and continue 
to hold the power button until the display 
indicates clr (about 3 seconds total) and then 
release the button. The stored maximum and 
minimum values are cleared. The unit returns 
to the normal mode.
The Min/Max memory may be configured to be automatically 
cleared when the unit shuts down, or may be configured to 
retain the values when the unit is shut off. See the section titled 
User Configuration Mode for these options.

Temperature Unit Selection

To change temperature units, press and hold 
the 

s

 button until the temperature indication 

is blanked and only the temperature units are 
displayed.
Then use the 

s

 and 

t

 buttons to select the 

desired temperature units. Standard units are 
Fahrenheit, Celsius, and kelvin. 
When the desired units are displayed, press 
and release the power button to save the 
selection and exit the change mode.
If no buttons are pressed for 15 seconds, the 
unit will automatically save the selection and 
exit the change mode.

Auto Shutoff Time Selection

To change the auto shutoff time, press and 
hold the 

t

 button until the auto shutoff time 

is displayed.
The lower display will indicate AST M if the 
auto shutoff time displayed is in minutes, and 
AST H if it is in hours. 
An auto shutoff time of zero signifies that the 
auto shutoff feature is disabled.
Use the 

s

 and 

t

 buttons to select 0, or 1, 

2, 5, 10, 15, 20, or 30 minutes, or 1, 2, 4, 
or 8 hours. 
When the desired auto shutoff time is 
displayed, press and release the power 
button to save the selection and exit the 
change mode.
If no buttons are pressed for 15 seconds, 
the unit will automatically save the displayed 
selection and exit the change mode.

PRTXB Series RTD Process Thermometers

2

User Configuration Mode

With the unit off, press and hold the 

s

 

button. 
Then press the power button. 
Release all buttons when the display 
indicates CFG.
Before the unit enters the configuration 
mode, the display initially indicates _ _ _ _  
with the first underscore blinking, and with 
CFGPC on the lower display. 
The unit will automatically revert to normal 
operation if no buttons are operated for 
approximately 15 seconds. To cancel and 
return to normal operation, press and release 
the power button without entering any pass 
code characters.
Enter the user-modifiable configuration pass 
code (3510 factory default).
Use the 

s

 and 

t

 buttons to set the left-

most digit to 3.
Press and release the power button to index 
to the next position. The 3 will remain, and 
the second position will be blinking.
Use the 

s

 and 

t

 buttons to select 5.

Press and release the power button to index 
to the next position. 3 5 will remain, and the 
third position will be blinking.
Use the 

s

 and 

t

 buttons to select 1.

Press and release the power button to index 
to the next position. 3 5 1 will remain, and 
the fourth position will be blinking.
Use the 

s

 and 

t

 buttons to select 0.

Press and release the power button to 
proceed with configuration procedures. Note: 
If an incorrect pass code is entered, the unit 
will return to the start of the pass code entry 
sequence.

MAX/MIN Capture Configuration

The upper display will be blank.
Use the 

s

 and 

t

 buttons to select from the 

following. 
MX/MN  Both highest and lowest values will 

be captured

MX/--  Only highest value will be captured
--/MN  Only lowest value will be captured
--/-- 

Capture feature is disabled

Press and release the power button to move 
on to the next parameter.

Auto/Manual MAX/MIN Clearing

The upper display will indicate clr.
Use the 

s

 and 

t

 buttons to select from the 

following.
AUTO  Maximum and minimum values will 

automatically be cleared whenever 
the unit shuts off.

MAN  Maximum and minimum values will 

be retained and must be cleared 
manually as desired.

Press and release the power button to save 
the user configuration and restart the unit.

ALARM 1

ALARM 2

DEG F

deg f

75.0

t

s

t

s

*

X X X X X

88888

ALARM 2

ALARM 1

deg f

off

t

s

Press any button to 

keep on

deg f

75.0

t

s

*

Press and hold 

to turn off

deg f

75.0

t

s

*

Press and hold 

about 1 sec.

 deg f

i25.5

t

s

  max

 deg f

i25.5

t

s

  min

*

Press until MIN 

is displayed

deg f

t

s

*

Press and hold 

about 1 sec.

 mx/mn

c l r

t

s

*

Press until clr 

 is displayed

normal mode

or

AST M

5

t

s

*

Press & hold

ast m

i0

t

s

ast h

*

or

*

AST H

t

s

*

Save & exit

normal mode

or

deg f

t

s

*

Press & 

hold

deg f

t

s

deg C

Kelvn

*

or

*

deg C

t

s

*

Save & exit

normal mode

cf6

t

s

Release buttons

t

s

*

2

*

1

cfgpc

____

t

s

cfgpc

3___

t

s

*

*

or

*

Increment 

up or down

Move to 

next #

cfgpc

35 i0

t

s

*

*

or

*

Increment 

up or down

Move to 

next #

cfgpc

35 i0

t

s

*

Press to enter con-

figuration mode

--/--

t

s

*

*

or

Select min 

max modes

MX/mn

t

s

*

Press to save and 

move to next

auto

c l r   

t

s

*

*

or

Select auto or 

manual clear

man

t

s

*

Press to save and 

restart

c l r   