PRTXB Series  

RTD Process Thermometers

M-4998/0818

MODEL

PROBE 

LENGTH

PROBE 

TYPE

PRTXB-2

2.5

"

Fixed

PRTXB-2-SL

2.5

"

Spring-Loaded

*

PRTXB-4

4

"

Fixed

PRTXB-4-SL

4

"

Spring-Loaded

*

PRTXB-6

6

"

Fixed

PRTXB-6-SL

6

"

Spring-Loaded

*

PRTXB-9

9

"

Fixed

PRTXB-9-SL

9

"

Spring-Loaded

*

PRTXB-12

12

"

Fixed

 PRTXB-12-SL

12

"

Spring-Loaded

*

*Thermowell required for spring-loaded versions

Range and Resolution

User selectable °F, °C or K 
  –58.0°F to 392.0°F 
  –50.0°C to 200.0°C 
  220.0K to 475.0K 
0.1 degree resolution

Typical Accuracy

Includes linearity error and ±1 LSD
11-point linearization  
±0.7°C at –50°C 
±0.4°C at 0°C 
±0.9°C at 100°C 
±1.4°C at 200°C

Display

4 readings per second nominal display update rate 
4 digit LCD, 0.5" H, 5 character 0.25" H alphanumeric 
BL models: red LED backlight

Sensor

IEC-751 Class B 100 

 Platinum RTD, 0.00385 alpha curve 

1

/

2

" NPT male, 316 stainless steel 

Spring-loaded probe versions fit standard thermowells
Fixed probe pressure rating: 5000 psi max. 
Fixed probes are welded to hex fitting

Auto Shutoff Time

Factory default 5 minutes. User settable to 1, 2, 5, 10, 15, 20, 
30 minutes, 1, 2, 4, 8 hours, or manual on/off.  
OFF warning before auto shutoff to allow reset of timer

Batteries, Battery Life, Low Battery Indication

B:   2 AA alkaline, approx. 1000 hours 
BL:   2 AA alkaline, approx. 150 to 750 hours depending on  
 

backlight usage.  

BL:   Button press activates backlighting for 1 minute 
Low battery symbol on display

Controls and Functions

Three front buttons for power on/off, min/max functions, 
selection of °F, °C, or K, auto shutoff times, calibration and, 
configuration options
User-defined pass codes for configuration and calibration to 
prevent unauthorized changes

Maximum and Minimum Readings

User-configurable maximum and/or minimum temperature 
indication. Factory default configuration MAX/MIN disabled.
Choice of MAX only, MIN only, MAX/MIN, or none
Option to retain or clear MAX/MIN temperatures at shutoff

Out-of-Range

ALARM1 under range indication on display 
ALARM2 over range indication on display 

Calibration

User settable pass code required to enter calibration mode 
Zero and span temperature calibration
Non-interactive zero, span, and linearity, ±10% of range

Weight

Product: 

12 ounces (approximately) 

Shipping: 

1 pound (approximately) 

Housing

ABS/polycarbonate NEMA 4X case, polycarbonate label, rubber 
rear gasket

Connection, Material, Media Compatibility

1/2" NPT male fitting, 316L stainless steel 
All wetted parts are 316L stainless steel 
Thermowell required for spring-loaded versions

Storage Temperature

–40 to 203°F (–40 to 95°C)

Operating Range

–4 to 185°F (–20 to 85°C) at housing

2.0"

Hex nipple  

1/2" NPT x 1/2" NPT 

 316 stainless

Spring-loaded models

approx. 0.25" longer 

to compress element 

in thermowell.

0.50" max.  

compression

1.25"

3.5"

0.75"

0.25" dia.

Stem 

Length  

L

Continuous weld 

on non-spring 

loaded models

1.37"

3.0"

X X X X X

88888

ALARM 2

ALARM 1

t

s

Measured from 

top of full threads 

to tip of probe. 

Length L = 

thermowell “A” or 

“element length”

Turn at 

hex 

fitting 

only!

Low battery 

indication

e-mail: [email protected] 

For latest product manuals: 

www.omegamanual.info

Shop online at 

omega.com

User’s Guide

WARNING:

 This product can expose you to chemicals including 

lead, nickel and chromium, which are known to the State of 

California to cause cancer or birth defects or other reproductive 

harm. For more information go to www.P65Warnings.ca.gov

TM

Отзывы: