Omega Engineering hot point CL900A Series Скачать руководство пользователя страница 14

3.3 

Changing the Temperature Setpoint

The layout of the front panel is shown in Figure 2. The CL900A/CCL950A
incorporates a PID digital setpoint controller. The setpoint variable (SV) is
changed by holding in either the 

key or the 

key. The setpoint variable

will change at 3 different rates depending on how long a key is being pressed.
The minimum setpoint is 38°C (100°F), the maximum setpoint 482.2°C (900°F).

3.4

Changing the Controller Parameter Settings

The CL900A/CL950A operates at its optimum performance when left with the
factory parameter settings. The only internal parameter that the operator may
need to change is the engineering units (°F or °C). Below are two diagrams: a)
menu hierarchy with factory default settings b) programming procedure. In the
event that some parameters have accidentally been changed, use the factory
default settings shown below.

NOTE: 

Changing any other parameter settings besides the engineering units will decrease the

accuracy of your calibrator.

Figure 8.  Controller Menu Hierarchy Showing Factory Default Parameter Settings

WARNING:

Do not re-program the controller’s parameter settings or change the wiring inside your unit to

override the maximum setpoint value of 482.2°C (900°F).

3

3-4

Operation

3

 

 

 

 

 

Menu 00 

Menu 01 

Menu 02 

Menu 03 

Menu 04 

Menu 05 

Key Lock 

SETPOINT 

Ac.Cd =  02 

Ac.Cd =  03 

Ac.Cd =  04 

Ac.Cd =  05

Ac.Cd          

Gn.o1

 113 

ALr1

 925 

id.no

 01 

SnSr 

P

Gr.o2

  

ALr2   

bAUd

  

12.o.7

Sn.00 

rAtE

  

24  

Cy.t1

  

01  

CAL.L   

dEC.P 

rSEt

  

122  

Cy.t2  

CAL.H   

FILt 

H.Hys

SP.tt

  

OFF    

OUt.1

  

Ht.P

HyS.1   

L.SP.L

80    

OUt.2

  

Air

C.HyS

L.SCL

CoL.t

  

nor

HyS.2   

U.SP.L

900    

A1.HL

  

HI

C.SPr

H.SCL

A1.Pd

  

Pr

SPr.2       

A1.OP

   LAt

dPnG

  

Nl      

A2.HL

  

Hl

A2.Pd

  

Pr

A2.OP

   OFF

Unit

  

F

Содержание hot point CL900A Series

Страница 1: ...www omega com e mail info omega com User sGuide NIST MADE IN Shop on line at CL900A CL950A SERIES hot point Dry Block Probe Calibrator TM TM...

Страница 2: ...Toll Free 0800 1 66342 e mail info omegashop cz France 11 rue Jacques Cartier 78280 Guyancourt France TEL 33 0 1 61 37 2900 FAX 33 0 1 30 57 5427 Toll Free in France 0800 466 342 e mail sales omega fr...

Страница 3: ...3 4 3 4 Changing the Controller Parameters 3 4 3 5 Target Plate Temperature Transition Chart 3 5 3 6 Replacing Probe Well Inserts in the CL900A 3 5 3 7 Fanning System 3 6 3 8 Overhead Reset Switch 3 7...

Страница 4: ...CL950A 230 Probe Well 2 1 6 Front Panel 3 1 7 Back Panel 3 3 8 Controller Menu Hierarchy Showing Factory Default Parameters Settings 3 4 9 Approximate Target Plate Temperature Transition Times Chart...

Страница 5: ...ain power cord before attempting any maintenance or fuse replacement Do not connect and or operate this unit to a non grounded non polarized outlet or power source Do not connect the serial port or re...

Страница 6: ...int and actual temperature are displayed simultaneously Temperature can be set in one degree increments from 22 C 40 F above ambient temperature to 482 C 900 F The CL900A differs from the CL950A only...

Страница 7: ...ndard metric holes with diameters 2 mm 3 mm 4 5 mm and 6 mm x 2 with 152 4 mm hole depth Figure 4 CL950A Probe Well Figure 5 CL950A M Probe Well Probe Well Inserts with Inch Sizes Probe Well Inserts w...

Страница 8: ...CL900A CL950A SERIES 1 1 4 NOTES...

Страница 9: ...n and operate at least ten inches from any air obstructions to the fan front panel rear panel bottom and top of the unit in an ambient environment between the specified 5 C to 38 C 40 F to 100 F 2 3 A...

Страница 10: ...ational 230 VAC 50 60 Hz models On CL900A 230 CL950A 230 models an International style power cord with the proper color code and approvals is provided with stripped wire ends for connection to the pro...

Страница 11: ...rol algorithm used in high end temperature controllers The controller causes the process to attain the desired temperature by turning the process on or off The process may be a heater or refrigerator...

Страница 12: ...r 3 seconds initially to access the programming mode Once in the programming mode momentarily press this key to scroll through the menus Raise Key Momentarily press to increase the selected parameter...

Страница 13: ...an will turn on immediately when entering the ON mode Once the calibrator is no longer being used and the switch is placed in the STANDBY mode the unit is powered down except for the fan circuit The f...

Страница 14: ...b programming procedure In the event that some parameters have accidentally been changed use the factory default settings shown below NOTE Changing any other parameter settings besides the engineerin...

Страница 15: ...rent probe well inserts available for the CL900A CL900A 230 The CL900A is shipped with the CL906 insert diameter 1 4 depth 6 and the CL900A 230 is shipped with the CL906 M insert diameter 4 5 mm depth...

Страница 16: ...f the shaft Pull the pin out slowly See Figure 6 Figure 10 Removal of Probe Well Insert in CL900A 3 7 Fanning System The CL900 950 has 2 cooling fans One is mounted just under the top panel The other...

Страница 17: ...one or both of the fans fail and the temperature inside the unit raise to an unacceptable level a thermal cutoff switch will disconnect the heater power The controller will still display the desired...

Страница 18: ...rom the unit 4 2 3 Fan and Vent Guards The fan guard and vent guards should be cleaned as a minimum annually A medium power vacuum cleaner with a brush attachment is recommended 4 3 Fuse replacement W...

Страница 19: ...e CL900A and the computer must be communicating with the same serial communications parameters to establish a working communication link The serial communications feature can be tested using terminal...

Страница 20: ...case R To read a parameter from the controller M To temporarily modify a controller parameter lost upon shutdown E To modify a controller param in non volatile mem saved even after shutdown parameter...

Страница 21: ...in from 38 C to 398 C from 100 F to 750 F Temperature Uniformity Within the Bottom 1 of the Test Well CL900A W 6 Well Inserts 0 3 rdg CL900A W 4 Well Inserts 0 4 rdg CL950A 0 3 rdg Max Probe Immersion...

Страница 22: ...nit requires service contact our customer service department d Unit has been operated above ambient temperature operating range Safety cutoff switch has disabled heater Unit should be sent back to fac...

Страница 23: ...chart so that its reading can be correlated to the actual value being measured IEC International Electrotechnical Commission NIST National Institute of Standards and Technology PID Proportional Integr...

Страница 24: ...ce Calibration Range 32 to 400 C 90 to 752 F Accuracy 1 4 C 2 5 F Emissivity 0 95 Ambient Temp 0 to 40 C 32 to 104 F Cavity Size 28 6 mm 1 125 in Power 115 230V 50 60 Hz 175W BB704 4 Target Plate Blac...

Страница 25: ...NOTES CL900A CL950A SERIES 9 2...

Страница 26: ...NOTES CL900A CL950A SERIES 9 3...

Страница 27: ...shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OMEGA...

Страница 28: ...city Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industria...

Отзывы: