background image

SPECIFICATIONS 

ELECTRICAL 

Temperature Scale:

 Celsius or Fahrenheit user-selectable 

Measurement Range: 

Pt385(100

) -200°C to 800°C, (-328°F to 1472°F) 

Pt3916/Pt3926(100

) -200°C to 630°C, (-328°F to 

1166°F) 

Resolution: 

0.1°C or 0.2°F 

Accuracy: 

Accuracy is specified for operating temperatures over the 
range of 18°C to 28°C (64°F to 82°F), for 1 year, not 
including RTD probe error. 
±(0.05% rdg + 0.2°C) on °C scale 
±(0.05% rdg + 0.4°F) on °F scale 

Temperature Coefficient: 

0.1 times the applicable accuracy specification per °C 
from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 
82°F to 122°F). 

Input Protection: 

24V dc or 24V ac rms maximum input voltage on any 
combination of input pins. 

Maximum Differential Common Mode Voltage 
(Maximum Voltage between T1 and T2 during meas-
urement): 

1volt. 

Reading Rate: 

1 time per second. 

Input Connector: 

Accepts for RTD 4 wires subminiature 4-Prong type 
connectors (flat blades spaced 7.9mm, center to center). 

 

Wireless Features: 

Frequency range: 910~920MHz 
Low current consumption less than 1mA 
The transmitting distance can reach 25M without mag-
netic interference. 
 

ENVIRONMENTAL 

Ambient Operating Ranges: 

0°C to 50°C (32°F to 122°F) <80% R.H. 

Storage Temperature: 

-20°C to 60°C (-4°F to 140°F) <70% R.H. 
 

GENERAL 

Display: 

4½ digit liquid crystal display (LCD) with maximum 
reading of 19999. 

Overload: 

“----.-” or “OL” is display. 

Battery: 

1.5V x 4 PCS (SIZE AAA) UM-4 R03. 

Battery Life: 

200 hours typical with carbon zinc battery. 

Auto power off: 

30 minutes, press power key to resume 

operation. 

Dimensions: 

160mm(H) x83mm(W) x 38mm(D). 

Weight: 

Approx. 255g including batteries. 

OPERATING INSTRUCTIONS 

1. “ ” Power Button 

The “

 

key turns the thermometer on or off. In the SET 

mode cannot be powered off. Exit SET mode to power 
off. 
 

APO function mode 

Press “

 

power key for more than 6 seconds to disable 

the auto-power function. The display will show “APO 
OFF”. 
 

2. °C/°F Selecting the Temperature Scale (Main 

display) 

Reading the main displayed in either degrees Celsius(°C) 
or degrees Fahrenheit(°F). When the thermometer is 
turned on, it is set to the temperature scale that was in use 
when the thermometer was last turned off. To change the 
temperature scale, press the “°C/°F” key. 
 

3. “ ” Display Back-Light 

Press the “ ” key to turn on or turn off the Back-Light. 
 

4. T1 T2/T1-T2 Main display Input Selection 

The input selection indicates which input is selected for 
main display; T1 probe, T2 probe or the difference be-
tween the two probes (T1-T2). When the thermometer is 
turned on, it is set to T1, when main display input se-
lected T1, then T1 input can select alternate of probe by 
pressing Pt385/Pt3926/Pt3916 key switch. 
 

5. TYPE(Pt385/Pt3926/Pt3916) Input RTD 

Probe Select (only Main display) 

The “TYPE” key switches the T1 input to select the 
Pt385, Pt3926 or Pt3916 RTD probe as input, when main 
display input selected T1. When the thermometer is 
turned on, it is set to the probe selected that was in use 
when the thermometer was last turned off. 
 

6. MIN/MAX with Time record mode 

(only Main display) 

Press “MIN/MAX” key to enter the MIN MAX Re-
cording mode, (displays the Maximum reading with time, 
Minimum reading with time and Average reading stored 
in record mode). In the this mode the automatic 
power-off feature is disabled and “ ”

 

key, “°C/°F” key, 

“REL” key, “SET” key, “Hi/Lo Limits” key and main 
display “T1 T2 T1-T2” key, “TYPE” key are disabled. 
The beeper emits a tone when a new minimum or maxi-
mum value is recorded. 
Press “MIN/MAX” key to cycle through the MAX, MIN 
and AVG readings. If an overload is recorded, the aver-
aging function is stopped. In this mode, press the 
“HOLD” key to stop the recording of readings, all values 
are held, press again to restart recording. 
To prevent accidental loss of MIN, MAX and AVG data, 
this mode can only be cancelled by pressing and holding 
the “MIN/MAX” key for 2 seconds. All recorded read-
ings are erased. 

7. REL Relative mode (only Main display) 

Press the “REL” key to enter the relative mode, zero the 
display, and store the displayed Reading as a reference 
value. REL is shown on the

 

display. Press “REL” key 

again to exit the relative mode. The relative reference 
value can also be entered by the user. (See “SET mode” 
later in this manual). When the desired relative value has 
been entered, press “REL” key to enter the relative mode 
and than press “SET” key use the entered relative value 
as a reference value. Press “REL” key again to exit the 
relative mode. In the relative mode, the value (can not 
>±1999.9 counts) shown on the LCD is always the dif-
ference between the stored reference and the present 
reading. 
 

8. HOLD mode (only Main display) 

Press the “HOLD” key to enter the Data Hold mode, the 
“HOLD” annunciator is displayed. When HOLD mode is 
selected, the thermometer held the present readings and 
stops all further measurements. Press the “HOLD” key 
again to cancel HOLD mode causing thermometer to 
resume taking measurements. In the MIN/MAX re-
cording mode, press “HOLD” key to stop the recording. 
Press “HOLD” key again to resume recording. (Previ-
ously recorded read are not erased). 
 

9. SET mode (Relative value set, Time set and 

Hi/Lo Limits value set)

 

9.1

 Press the “SET” key to enter relative values SET 

mode (Press “ENTER” key   to  skip  setting  relative 
value). = = = =.= is displayed on the main display. 
The relative value is entered via overlay numbers, 
press overlay “ENTER” key to store the relative value, 
and advance to elapsed time set mode. 

9.2

 Elapsed time set mode, (Press ENTER key to skip 

Elapsed time set mode) =.= = = : = = is displayed in 
second and third display. Time (hours, minutes, sec-
onds) value is entered via overlay numbers, press 
overlay “ENTER” key to store time value. Elapsed 
time starts from set time value. 

9.3

 Hi Limit value set mode, (Press “ENTER” key to skip 

Hi Limit value set mode), = = = =.= is displayed in 
main display, Hi Limit value is entered via overlay 
numbers, then press “ENTER” key to store the Hi 
Limit value. = = = =.= is displayed in main display, 
Lo Limit value is entered via overlay numbers, then 
press overlay “ENTER” key to store the Lo Limit 
value and exit SET mode. 

9.4

 When the thermometer is turned on it uses the relative 

value and Hi/Lo Limits values that were entered when 
thermometer was last in use. 

 

10. T1/T2 T1-T2 second display Input Selection 

The input selection indicates which input is selected for 
second display; T1 probe, T2 probe or the difference 
between the two probes (T1-T2), when the thermometer 
is turned on, it is set to T2 input can select alternate of 
probe by second display Pt385/Pt3926/Pt3916 key 
switch. 

11. TYPE(Pt385/Pt3926/Pt3916) Input RTD 

Probe select (second display) 

The Pt385/Pt3926 key switches the T2 input to select the 
Pt385, Pt3926 or Pt3916 RTD probe as input, when sec-
ond display input selected T2. When the thermometer is 
turned on, it is set to the RTD selected that was in use 
when the thermometer was last turned off. 
 

12. Hi/Lo Limits mode (only Main display) 

Press “Hi/Lo Limits” key to enter the Hi/Lo Limits com-
parative mode. When the input temperature value ex-
ceeds the Hi or Lo Limits value, the beeper emits a con-
tinuous or pulsed tone. Press “Hi/Lo Limits” key again to 
exit the Hi/Lo Limits mode. 
 

13. WIRELESS MODE: 

Press the “ ” key for more than two seconds to start 
wireless function. Press the “ ” key again for another 
two seconds to stop wireless function. The wireless mode 
will shut down if there is no wireless signal for two min-
utes. 
To SET CH/ID to 00,00, press the “Hi/Lo Limits” key 
and “

” power key for more than 6 seconds with the 

meter powered down. The meter will set channel and ID 
to 00,00 status. The second display will show 00, which 
means that the channel and ID has been set to 00. 

To check the channel and ID of the meter:

 

When the meter is off, press “°C/°F” key and “

” for 5 

seconds, LCD’s main display will show channel number, 
the second display will show ID number. 