background image

5

GP901    M-3343  05/08

Gauge Head Application Information

When mounting the Gauge Head in a fixture, care must be taken as the linear bearing assembly at the 

tip end is made to very fine tolerances. If a high force is applied to the probe casing performance may be 

affected. Ideally probes should be clamped in a yoke, split or collar clamp. If single point screw clamping 

is adopted (see diagram), then a tightening torque (in Nm) should be limited to that which will give a 

maximum of 50Kg point load. The Gauge Head may be mounted in any attitude.

 

                                                          where 

d = Screw dia. (mm) 

 

                                                                    

P = Screw pitch (mm)

 

                                           

 (assumes a V form thread and 0.15 coefficient of friction).

Typical maximum torque is 0.27 Nm (i.e. M5 x 0.8 steel or 10-32 UNF steel)

A clearance hole in the fixturing of 9.5 mm 

F

 is desirable around the gaiter for satisfactory operation.

Tip

Generally, where piece parts are automatically loaded into a fixture, or where there is the possibility of 

severe side loads, the use of pneumatically operated transducers is strongly recommended.

To remove the tip, the shaft should be fully retracted into the body before unscrewing.

Pneumatic Gage Heads

The tip of a pneumatic Gage Head is normally biased in the inward direction and the extension of the tip is 

achieved by applying air pressure to the Gauge Head nozzle.

Air pressure may be varied within the specified limits to achieve the desired tip force at zero. The 

pneumatic Gage Heads are designed to be leak free and of small volume to minimise the air flow 

requirements and maintain a consistent tip force. Care must be taken not to damage the gaiter seal as the 

satisfactory operation of the Gage Head is entirely dependent on this component. A clearance hole in the 

fixturing of 9.5 

mm 

F

 is desirable around the gaiter for satisfactory operation.

To maximise working life of the Gage Heads the air supply should be both clean and dry for continual 

reliable operation. Maximum relative humidity of 60 % RH and filtered to better than 5 µm particle size.

Environmental Considerations

Most Gage Heads are designed to be used with a gaiter and are designed to withstand the rigours of 

in-process gauging and will operate satisfactorily in the presence of copious amounts of coolants and 

cleaning solvents.

For Gage Heads without a gaiter such as Feather touch or in situations where removal of the gaiter is 

necessary (i.e. at low temperatures where stiffening of the gaiter may be a problem) then the working 

atmosphere must be both clean and dry. Ingress of dirt or fluids may interfere with the performance of the 

probe or even cause damage.

Содержание GP901 Series

Страница 1: ...Series 956 User s Guide http www omega com e mail info omega com GP901 Series Digital Probes...

Страница 2: ...69 6311114 Toll Free 0800 1 66342 e mail czech omega com France 9 rue Denis Papin 78190 Trappes Tel 33 130 621 400 FAX 33 130 699 120 Toll Free in France 0800 4 06342 e mail france omega com Germany A...

Страница 3: ...first are the specifications for the complete Digital Probe PIE gauge head and the second is mechanical data relevant to the gauge head only PIE is designed to be primarily used with the T CON connect...

Страница 4: ...S 6 0 05 mm 6 0 2 x D 1 mm Probes i e GP901 1 S 6 0 1 mm 6 0 1 x D 2 mm Probes i e GP901 2 S GP901 2 P 6 0 1 mm 6 0 1 x D 5 mm Probes i e GP901 5 S GP901 5 P 6 0 1 mm 6 0 2 x D 10 mm Probes i e GP901...

Страница 5: ...ward direction and the extension of the tip is achieved by applying air pressure to the Gauge Head nozzle Air pressure may be varied within the specified limits to achieve the desired tip force at zer...

Страница 6: ...lity flex life chemical resistance abrasion resistance electrical and electrical screening To minimise transducer failure due to cable damage cable runs should be positioned well clear of moving compo...

Страница 7: ...GP901 M 3343 05 08 PIE and T CON Installation...

Страница 8: ...GP901 M 3343 05 08 This Page Intentionally Left Blank...

Страница 9: ...GP901 M 3343 05 08 This Page Intentionally Left Blank...

Страница 10: ...10 GP901 M 3343 05 08 This Page Intentionally Left Blank...

Страница 11: ...otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sol...

Страница 12: ...cators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Cond...

Отзывы: