Close Up of Label Showing Flow Direction 

(FLR 1000 shown, other models similar) 

 

 

 

The tubing ID should be as large as possible to avoid restrictions and 
turbulence. For the best results, a straight length of tubing before the unit 

is recommended. This should be 20 times the tubing I.D. or greater  

 

 

4.

 

Electrical Connections 

 

 

Caution: Incorrect wiring may cause severe damage to the unit. 
Applying an AC voltage (115VAC or 230VAC) directly to the unit 

will cause damage. Read the following instructions carefully 
before making any connections.  

 

 

 

a)

 

Overview 

 

Omega Engineering flow sensors for gases provide a 0-5VDC output 

proportional to the volumetric flow rate. This output may be connected to 

a display, data acquisition system or voltmeter. The FLR 1000-D and 1000-
BR-D series feature an integral display that provides a local flow reading. 

These units also have a 0-5VDC analog output available. If required, this 
may be connected to another display, data acquisition system or 

voltmeter. 
 

A stable D.C. power supply is required to operate the unit. The voltage 

and current requirements depend on the configuration of the unit. Full 

details may be found in the Specification section of this manual. 
 

Connecting wires should be as short as possible to avoid voltage drops. 
Twisted 2 pair conductor cable of a suitable gauge should be used if the 

length of the power wires is to be longer than 1 meter (3.2 feet). 

 

Содержание FLR 1000 Series

Страница 1: ...FLR 1000 1000 BR Series Flow Sensors Meters For Gases...

Страница 2: ......

Страница 3: ...onnections Voltage Output Units 10 d Using a 0 5VDC Output Power Adapter Package 11 C Operation 12 1 Start Up 12 2 Flow Readings 12 a 0 5VDC Analog Outputs 12 b Units with an Integral Display 12 3 Ope...

Страница 4: ...al Contents of Box Sensor Calibration Certificate Instruction Manual Model FLR 1000 shown other models similar Caution Take care not to drop your sensor or subject it to sudden impact Read the install...

Страница 5: ...eam is reflected off each white section The reflected beam is detected by a phototransistor that converts the reflections into electrical pulses As the wheel spins faster the pulse rate increases Proc...

Страница 6: ...lled upstream of before the unit 2 Mounting the Flow Sensor or Flow Meter Caution On high flow units 2 0 10 0 L min and above DO NOT disassemble the shunt mounting plate assembly or the unit will not...

Страница 7: ...the unit may be effected and leaking may occur If different fittings are required please contact the Omega Engineering Company Service Department for assistance When connecting the sensor to the tubi...

Страница 8: ...e a 0 5VDC output proportional to the volumetric flow rate This output may be connected to a display data acquisition system or voltmeter The FLR 1000 D and 1000 BR D series feature an integral displa...

Страница 9: ...lor cross reference may also be found in Appendix F Page 18 of this manual b Connecting a Cable Assembly The connector on the end of the cable assembly should be pushed into the mating socket on the s...

Страница 10: ...e should be connected to the Positive of the power source The BLACK wire should be connected to the Negative Ground of the power source The WHITE wire provides the signal output and should be connecte...

Страница 11: ...utput may be connected to a display data acquisition system or voltmeter with a minimum load of 2 5 k kilo ohms The YELLOW wire of the cable assembly provides the 0 5VDC signal output and should be co...

Страница 12: ...ibrated using AIR as the reference media unless otherwise noted Calibrations are carried out with the mounting feet horizontal on a flat horizontal surface a 0 5VDC Analog Outputs By monitoring the vo...

Страница 13: ...rranty Flow measurement is only accurate within the calibrated flow range for the unit This is detailed on the calibration certificate The unit will still operate to some degree outside this flow rang...

Страница 14: ...g other gases will effect the calibration The amount of calibration error will depend on the characteristics of the gas being flowed The flow reading correction factor or calibration factor may be app...

Страница 15: ...brated for the new gas For example A 2 0 10 0 l min unit that has been calibrated for Air is to be used with CO2 The correction factor calculated as above is 0 88 When flowing CO2 the unit will read h...

Страница 16: ...s sealed 2 Cleaning and Flushing If there is a build up of deposits or particles from the measured gases it may be necessary to clean or flush the unit This should be done by flowing clean particle fr...

Страница 17: ...itting Material Standard Acetal Brass Fitting Material Optional Acetal brass or stainless steel Recommended Filtration 25 microns or less Compatible Gases Clean dry 0 5VDC Output Signal Standard Minim...

Страница 18: ...or Pin And Wire Color Cross Reference Pin Configuration for Connector Socket and Connector Pin Cable Wire Color Description 1 Black Signal Power Negative Ground 2 White Voltage Output 3 Red Power Posi...

Страница 19: ...pg 19 of 29 G Dimensions ALL DIMENSIONS IN INCHES MILLIMETERS IN BRACKETS FLR 1000 For Gases 20 200 ml min To 1 0 5 0 l min Units 1 4 Acetal Fittings Shown...

Страница 20: ...pg 20 of 29 FLR 1000 For Gases 2 0 10 0 l min Units 1 4 Acetal Fittings Shown...

Страница 21: ...pg 21 of 29 FLR 1000 For Gases 4 0 20 0 l min To 100 500 l min Units 1 2 Acetal Fittings Shown...

Страница 22: ...pg 22 of 29 FLR 1000 D For Gases 20 200 ml min To 1 0 5 0 l min Units 1 4 Acetal Fittings Shown...

Страница 23: ...pg 23 of 29 FLR 1000 D For Gases 2 0 10 0 l min Units 1 4 Acetal Fittings Shown...

Страница 24: ...pg 24 of 29 FLR 1000 D For Gases 4 0 20 0 l min To 100 500 l min Units 1 2 Acetal Fittings Shown...

Страница 25: ...pg 25 of 29 FLR 1000 BR D For Gases 20 200 ml min To 1 0 5 0 l min Units 1 4 Brass Fittings Shown...

Страница 26: ...tion B5 and ensure connector is correctly plugged in Allow unit to dry out in a dry environment Unit must be returned for repair see Section J Flow indication bounces or fluctuates Flow is actually fl...

Страница 27: ...ection J Flow reading accuracy is outside the specification Gas has a specific gravity other than Air at 20 C Moisture in sensor Mechanical sensor assembly damaged Pressure drop issue Calculate the fl...

Страница 28: ...pg 28 of 29...

Страница 29: ...M4575 0807 pg 29 of 29...

Отзывы: