DARKFIELD PROBLEMS 

Problem

 

Cause

 

Solution

 

It’s totally dark in 
the viewing field 

The light is not on 

Turn on the light 

(A191BOIL only)

 

There is no oil in 

between the condenser top lens 
and slide

 

Place a drop of oil on the top lens of 
condenser and let it contact the underside 
of slide 

(A191BOIL only)

 

The built-in iris 

diaphragm of 100X objective is not 
in proper position

 

Adjust the iris diaphragm ring on the 
objective 

The condenser is not in the right 
position 

Lower or raise the condenser slightly to the 
position that the specimen is the brightest. 

 

 

(A191BOIL only)

 

Make sure the oil contact 

the bottom of the slide and the top of the 
condenser during the adjustment.

 

The illumination is 
insufficient or too 
bright 

The intensity of lamp is too low 

Adjust the intensity dial to increase the 
brightness 

The condenser is not in the right 
position 

Lower or raise the condenser slightly to the 
position that the specimen is the brightest.   

 

(A191BOIL only)

 

Make sure the oil contact 

the bottom of the slide and the top of the 
condenser during the adjustment.

 

The condenser is not centered 
properly 

Adjust the translational centering screws to 
center the condenser 

(A191BOIL only)

 

The built-in iris 

diaphragm of 100X objective is not 
in proper position

 

Turn the iris diaphragm ring on the 
objective 

Image of the 
specimen is not 
clear and lacking in 
sufficient contrast 

The specimen is not suitable for 
darkfield observation 

Change to brightfield or phase contrast 

(A191BOIL only)

 The built-in iris 

diaphragm of 100X objective is not 
in proper position 

Turn the iris diaphragm ring on the 
objective 

The condenser is not in the right 
position 

Lower or raise the condenser slightly to the 
position that the specimen is the brightest.   

 

(A191BOIL only)

 Make sure the oil contact 

the bottom of the slide and the top of the 
condenser during the adjustment. 

 

 

Содержание M834 Series

Страница 1: ...User Manual Trinocular Compound Microscope Model M834 Series MicroscopeNet com...

Страница 2: ...1 Component Illustration 3 2 Installation 4 3 Operation 7 4 Specifications 10 5 Optional Parts 11 6 Troubleshooting Guide 13 7 Darkfield Condenser Installation and Operation Instructions 15 8 Phase C...

Страница 3: ...l electrical connectors power cord should be inserted into an electrical surge suppressor to prevent damage due to voltage fluctuations 6 Important the lamp lamphouse and adjacent parts will become ve...

Страница 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Страница 5: ...11 Microscope Base 19 Condenser Holder 4 Slide Holder 12 Eyepiece 20 X Y Stage Moving Knobs 5 Condenser Lock Thumb Screw 13 Diopter Adjusting Ring 21 Light Collector 6 Condenser Focus Knob 14 Eyepiec...

Страница 6: ...iece tubes 2 Insert the eyepieces into the eyepiece tubes 2 4 Installation of the objectives 1 Adjust the coarse focus knob until the mechanical stage is at its lowest position 2 Turn the caps counter...

Страница 7: ...ight housing 3 Take the eyepieces off and check if the view head is securely locked by the set screw Put the microscope down on its back gently 4 Open the cover on the bottom of the base by pull the k...

Страница 8: ...oto tube and then connect the camera to a computer via USB2 0 cable See Fig 6 3 The manual for the camera is on a CD or mini CD Refer to the manual to install the driver and software on to the compute...

Страница 9: ...il the specimen is close to the objective 2 Turn the coarse focus knob until the specimen is in focus 3 Use the fine focus knob to obtain a sharp image 4 To get a good focused image you may need to co...

Страница 10: ...sting condenser 1 Turn the condenser focus knob to raise or lower the condenser 2 The condenser is raised when using high magnification objectives and lowered when using low magnification objectives N...

Страница 11: ...microscope into focus by following the procedures in 3 3 2 Install the camera by following the procedures in 2 8 3 Open image observing software to examine more details see camera s manual 4 If the im...

Страница 12: ...Coaxial coarse and fine focusing knobs on both sides w stage upward stopper Minimum fine focusing adjustment at 0 002mm range 30mm Mechanical Stage Large double layer mechanical stage Dimension 5 1 2...

Страница 13: ...S Windows 32 64 bit Mac OS Included A1510 1280 x 1024 1 3MP A1520C 1600 x 1200 2 0MP A1530X 2048 x 1536 3 0MP A1550X 2592 x 1944 5 0MP A1590 3488 x 2616 9 0MP MS Windows 32 64 bit 2 Darkfield Condense...

Страница 14: ...Achromatic 20X with built in phase plate Achromatic 40X with built in phase plate spring Achromatic 100X with built in phase plate spring oil five positions 10 for 10X phase contrast objective 20 for...

Страница 15: ...17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Imm...

Страница 16: ...e condenser is not in the right position Lower or raise the condenser slightly to the position that the specimen is the brightest A191BOIL only Make sure the oil contact the bottom of the slide and th...

Страница 17: ...e condenser up or down slightly to get the best darkfield viewing Note The dry darkfield condenser is used with the dry objectives only The dry darkfield condenser works with the 4X 10X 40X objectives...

Страница 18: ...objectives a Take off all the objectives from the nosepiece b Install the phase contrast objectives onto the nosepiece following the steps in 2 4 2 Mounting the phase contrast condenser annular ring d...

Страница 19: ...on focusing etc following the instructions in this manual Note The phase contrast condenser will be working as a conventional Abbe condenser if the annular ring disk being put at B position 8 2 Phase...

Страница 20: ...he telescope The bright ring and dark ring should be coincided with each other as shown in Fig 21 d e If the ring images are not clear turn the top of telescope until both ring images are in focus f I...

Страница 21: ...is hard to find put a normal white print paper on the stage and under the objective then you will see the dark ring as shown in Fig 24 a h If the bright ring is still obscure as in Fig 24 b raise or l...

Отзывы: