Troubleshooting Guide 

GENERAL PROBLEMS

 

Problem

 

Cause

 

Solution

 

Lamp does not light 

when switched on

 

No electrical power   

Check power cord connection   

Lamp bulb burnt out   

Replace bulb   

Fuse blown out 

Replace fuse 

Darkness at the 
periphery or uneven 
brightness in the field 
of view 

Revolving nosepiece not in click stop 
position 

Revolve the nosepiece to click-stop 
position by swinging the objective 
correctly into the optical path 

The light source of the bulb is not at the 
center 

Adjust the position of the bulb 

Dirt or dust on the 
view 

Dirt or dust on the eyepiece, condenser, 
objective or collector lens or specimen 

Clean the lens with a lens cleaning 
paper 

Poor image quality or 
not able to get 
focused image 

No slide cover attached to the slide 

Attach a 0.17mm slide cover 

Slide cover is too thick or thin 

Use a slide cover of the appropriate 
thickness (0.17mm) 

Slide may be upside down 
(specimen at the bottom) 

Turn slide over so the cover-glass faces 
up 

Immersion oil is on a dry objective 
(especially the 40X) 

Check the objectives, clean if necessary 

No immersion oil used with 100X 
objective 

Use immersion oil 

Air bubbles in immersion oil 

Remove bubbles 

Condenser aperture is closed or open 
too much 

Open or close properly 

Condenser is positioned too low 

Position the condenser upward 

Specimen rises from stage surface 

Secure the specimen in the slide holder 

Blue filter not used 

Use daylight blue filter 

Lamp intensity is too high or low 

Adjust the light intensity by rotating the 
intensity control dial 

Slippage of focus 
when using the coarse 
focusing knob Fine 
focus is ineffective 

Tension adjustment is set too low 

Increase the tension on the focusing 
knobs 

Tension adjustment is set too high 

Loosen the tension on the focusing 
knobs 

Содержание M834 Series

Страница 1: ...User Manual Trinocular Compound Microscope Model M834 Series MicroscopeNet com...

Страница 2: ...1 Component Illustration 3 2 Installation 4 3 Operation 7 4 Specifications 10 5 Optional Parts 11 6 Troubleshooting Guide 13 7 Darkfield Condenser Installation and Operation Instructions 15 8 Phase C...

Страница 3: ...l electrical connectors power cord should be inserted into an electrical surge suppressor to prevent damage due to voltage fluctuations 6 Important the lamp lamphouse and adjacent parts will become ve...

Страница 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Страница 5: ...11 Microscope Base 19 Condenser Holder 4 Slide Holder 12 Eyepiece 20 X Y Stage Moving Knobs 5 Condenser Lock Thumb Screw 13 Diopter Adjusting Ring 21 Light Collector 6 Condenser Focus Knob 14 Eyepiec...

Страница 6: ...iece tubes 2 Insert the eyepieces into the eyepiece tubes 2 4 Installation of the objectives 1 Adjust the coarse focus knob until the mechanical stage is at its lowest position 2 Turn the caps counter...

Страница 7: ...ight housing 3 Take the eyepieces off and check if the view head is securely locked by the set screw Put the microscope down on its back gently 4 Open the cover on the bottom of the base by pull the k...

Страница 8: ...oto tube and then connect the camera to a computer via USB2 0 cable See Fig 6 3 The manual for the camera is on a CD or mini CD Refer to the manual to install the driver and software on to the compute...

Страница 9: ...il the specimen is close to the objective 2 Turn the coarse focus knob until the specimen is in focus 3 Use the fine focus knob to obtain a sharp image 4 To get a good focused image you may need to co...

Страница 10: ...sting condenser 1 Turn the condenser focus knob to raise or lower the condenser 2 The condenser is raised when using high magnification objectives and lowered when using low magnification objectives N...

Страница 11: ...microscope into focus by following the procedures in 3 3 2 Install the camera by following the procedures in 2 8 3 Open image observing software to examine more details see camera s manual 4 If the im...

Страница 12: ...Coaxial coarse and fine focusing knobs on both sides w stage upward stopper Minimum fine focusing adjustment at 0 002mm range 30mm Mechanical Stage Large double layer mechanical stage Dimension 5 1 2...

Страница 13: ...S Windows 32 64 bit Mac OS Included A1510 1280 x 1024 1 3MP A1520C 1600 x 1200 2 0MP A1530X 2048 x 1536 3 0MP A1550X 2592 x 1944 5 0MP A1590 3488 x 2616 9 0MP MS Windows 32 64 bit 2 Darkfield Condense...

Страница 14: ...Achromatic 20X with built in phase plate Achromatic 40X with built in phase plate spring Achromatic 100X with built in phase plate spring oil five positions 10 for 10X phase contrast objective 20 for...

Страница 15: ...17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Imm...

Страница 16: ...e condenser is not in the right position Lower or raise the condenser slightly to the position that the specimen is the brightest A191BOIL only Make sure the oil contact the bottom of the slide and th...

Страница 17: ...e condenser up or down slightly to get the best darkfield viewing Note The dry darkfield condenser is used with the dry objectives only The dry darkfield condenser works with the 4X 10X 40X objectives...

Страница 18: ...objectives a Take off all the objectives from the nosepiece b Install the phase contrast objectives onto the nosepiece following the steps in 2 4 2 Mounting the phase contrast condenser annular ring d...

Страница 19: ...on focusing etc following the instructions in this manual Note The phase contrast condenser will be working as a conventional Abbe condenser if the annular ring disk being put at B position 8 2 Phase...

Страница 20: ...he telescope The bright ring and dark ring should be coincided with each other as shown in Fig 21 d e If the ring images are not clear turn the top of telescope until both ring images are in focus f I...

Страница 21: ...is hard to find put a normal white print paper on the stage and under the objective then you will see the dark ring as shown in Fig 24 a h If the bright ring is still obscure as in Fig 24 b raise or l...

Отзывы: