Darkfield Condenser Installation and Operation 
Instructions 

7.1  Dry darkfield condenser of A191-INF 

1)  Mounting the dry darkfield condenser 

  Loosen the condenser lock thumb screw on the condenser holder and remove the 

brightfield condenser as shown in 

Fig. 11 (a)

  Install the dry darkfield condenser and tighten the condenser lock thumb screw on 

the condenser holder as shown in 

Fig. 11 (c)

 

 

 

2)  Centering the dry darkfield condenser 

  Turn the 4X objective to the light path.  

  Turn the condenser focus knob to lower the 

condenser till a dark spot showed in the viewing field 
as shown in 

Fig. 12

 

(a)

  Turn the condenser translational centering screws to 

move the dark spot to the center as shown in 

Fig. 12

 

(b)

3)  Place the slide on the stage.  

4)  Raise the condenser all the way to the top and lower it a little bit.  

5)  Following the procedures in this manual to focus and observe. 

6)  Move the condenser up or down slightly to get the best darkfield viewing. 

 
Note

 

  The dry darkfield condenser is used with the dry objectives only. 

  The dry darkfield condenser works with the 4X, 10X, 40X objectives. 

 

The dry darkfield condenser won

’t work with the 100X oil immersion objective.

 

 
 
 
 
 
 
 

  

 

 

 

 

 

(a)                (b) 

Fig.11 

(a) 

(b) 

(c) 

Fig.12 

(a) 

(b) 

Condenser Centering Screws 

(C) 

Condenser Lock Thumb Screw 

 

(c) 

Condenser Holder 

 

Содержание M8333S

Страница 1: ...User Manual Trinocular Infinity Compound LED Microscope Model M8333S series MicroscopeNet com...

Страница 2: ...1 Components Illustration 3 2 Installation 5 3 Operation 7 4 Specifications 10 5 Optional Parts 11 6 Troubleshooting Guide 13 7 Darkfield Condenser Installation and Operation Instructions 15 8 Phase...

Страница 3: ...is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto the stage objective or any other component disconnect the power cord immediately and wipe up the...

Страница 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Страница 5: ...www microscopenet com 3 1 Components Illustration 11 12 13 14 15 16 17 7 18 7 1 2 3 4 5 6 7 8 9 10 19 7...

Страница 6: ...4 Nosepiece 14 Objectives 24 Condenser 5 Slide Holder 15 Microscope Body 25 Aperture Diaphragm Ring 6 Mechanical Stage 16 Brightness Intensity Dial 26 Condenser Centering Screws 7 Light Collector 17...

Страница 7: ...sure the viewing head is installed securely 2 2 Installation of the photo tube 1 Remove the cap on the top of the viewing head 2 Thread the photo tube into the top of the viewing head 2 3 Installation...

Страница 8: ...The color filter is simply put into the filter holder 2 6 Connecting the power cord 1 Turn the power switch to the off position 2 Connect the power cord to the socket at the back of microscope body an...

Страница 9: ...e center of viewing field Caution Be sure not to allow an objective to touch a specimen slide when changing objectives 3 3 Focusing 1 With the 10X objective in position raise the mechanical stage usin...

Страница 10: ...3 5 Adjusting eyepiece diopter 1 Using the 10X objective and your right eye only observe your specimen through the eyepiece and bring it into focus by adjusting the focus knobs 2 Then observe the spe...

Страница 11: ...ive clean the 100X objective lens first 3 7 Adjusting iris aperture diaphragm Turn the aperture diaphragm ring to adjust the aperture size 3 8 Adjusting focus knob tension The tightness of the focus k...

Страница 12: ...e Double layer mechanical stage Dimension 5 1 2 x 4 1 4 140mm x 110mm Translation range 3 x 1 13 16 78mm x 46mm Condenser NA 0 9 0 25 with a flip up top lens built in aperture iris diaphragm Center ad...

Страница 13: ...ion Operating System Software A35100U CMOS USB2 0 3584 x 2748 10 MP MS Windows Mac OS 10 8 and up Linux 2 6 or above Included A35140U 4096 x 3288 14 0MP A3550U3 CMOS USB3 0 2560 x 1922 5 0MP A3580U3 3...

Страница 14: ...ntering Telescope A1PHIP Plan achromatic 10X with built in phase plate Plan achromatic 20X with built in phase plate Plan achromatic 40X with built in phase plate spring Plan achromatic 100X with buil...

Страница 15: ...en Clean the lens with a lens cleaning paper Poor image quality or not able to get focused image No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a...

Страница 16: ...piece is not in the click stop position Revolve the nosepiece to the click stop position Lamp intensity is too high or low Adjust the light intensity by rotating the intensity control dial Slippage of...

Страница 17: ...condenser till a dark spot showed in the viewing field as shown in Fig 12 a Turn the condenser translational centering screws to move the dark spot to the center as shown in Fig 12 b 3 Place the slide...

Страница 18: ...owing in the viewing field Raise the condenser until the darkfield image showing in the viewing field 4 When using the 100X objective to observe the specimen Raise the condenser until the top is close...

Страница 19: ...Mounting the phase contrast objectives a Take off all the objectives from the nosepiece b Install the phase contrast objectives onto the nosepiece following the steps in 2 4 2 Mounting the phase cont...

Страница 20: ...19 to move the dark spot to the center until two ring images are coincided or Turn the adjustment knobs Fig 19 to move the dark spot to the center until two ring images are coincided f Remove the cent...

Отзывы: