Troubleshooting Guide 

GENERAL PROBLEMS

 

Problem

 

Cause

 

Solution

 

Lamp does not 
light when 
switched on 

No electrical power   

Check power cord connection   

LED or power unit dead 

Replace LED light

 

Fuse blown out 

Replace fuse 

Darkness at the 
periphery or 
uneven 
brightness in the 
field of view 

Revolving nosepiece not in click 
stop position 

Revolve the nosepiece to click-stop 
position by swinging the objective 
correctly into the optical path 

Dirt or dust on 
the view 

Dirt or dust on the eyepiece, 
condenser, objective, collector lens 
or specimen 

Clean the lens with a lens cleaning 
paper 

Poor image 
quality or not 
able to get 
focused image 

No slide cover attached to the slide  Attach a 0.17mm slide cover 

Slide cover is too thick or thin 

Use a slide cover of the appropriate 
thickness (0.17mm) 

Slide may be upside down 
(specimen at the bottom)

Turn slide over so the cover-glass faces 
up

Diopter adjustment is not set 
properly 

Readjust the diopter settings 

Condenser aperture is closed or 
open too much 

Open or close properly 

Condenser is positioned too low 

Position the condenser upward 

Specimen rises from stage surface 

Secure the specimen in the slide holder 

Revolving nosepiece is not in the 
click-stop position

Revolve the nosepiece to the click-stop 
position

Lamp intensity is too high or low 

Adjust the light intensity by rotating the 
intensity control dial 

Slippage of focus 
when using the 
coarse focusing 
knob Fine focus 
is ineffective 

Tension adjustment is set too low 

Increase the tension on the focusing 
knobs

Tension adjustment is set too high 

Loosen the tension on the focusing 
knobs 

 

Содержание M8311E Series

Страница 1: ...User Manual Digital Compound Binocular LED Microscope Model MD8211E30 MicroscopeNet com ...

Страница 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 1 Components Illustration 3 2 Installation 4 3 Operation 6 4 Specifications 8 5 Troubleshooting Guide 9 ...

Страница 3: ...urface 4 If any specimen solutions or other liquids splash onto the stage objective or any other component disconnect the power cord immediately and wipe up the spillage Otherwise the instrument may be damaged 5 All electrical connectors power cord should be inserted into an electrical surge suppressor to prevent damage due to voltage fluctuations 6 For safety when replacing the LED light and the ...

Страница 4: ...cally using an air stream from an air bulb If dirt remains on the optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of absorbent cotton wound on the end of a tapered stick makes a useful tool for cleaning recessed optical surfaces Avoid using an excessive amou...

Страница 5: ...obs 3 Eyepiece Tube 10 USB Cable 17 Power Switch 4 Viewing Head w Built in Camera 11 Head Lock Thumb Screw 18 Brightness Intensity Dial 5 Objective 12 Nosepiece 19 Condenser Lock Thumb Screw 6 Slide Holder 13 Microscope Body 20 Condenser Focus Knob 7 Mechanical Stage 14 Coarse Focus Knob 21 Condenser 1 10 2 3 5 6 7 8 11 12 13 14 15 16 17 18 19 20 21 9 4 ...

Страница 6: ...e plastic cases and turn each one clock wise into the holes on the nosepiece Install the 4X objective into the nosepiece first Then in a counter clockwise direction rotate the nosepiece and install each succeeding higher magnification objective Note Inspect the objectives frequently for dirt or oil clean if necessary Use the 10X objective to initially focus the image of your specimen When changing...

Страница 7: ...ee Fig 3 a 3 Open the board and find the LED lamp See Fig 3 b 4 Loosen the 2 screws and unplug the LED connector 5 Replace the dead LED lamp with a new one 6 Tighten the 2 screws and plug in the LED connector 7 Put the board back its place 8 Tighten the 7 screws Fig 2 Color Filter Color Filter Holder Fig 3 7 Screws 2 Screws LED Lamp LED Connector a b c ...

Страница 8: ...ecimen is close to the objective 2 Turn the coarse focus knob until the specimen is in focus Then use the fine focus knob to obtain a sharp image You may now switch to another magnification objective 3 To get a good focused image you may need to combine the focus knob adjustment and interpupillary distance adjustment along with eyepiece diopter adjustment stated in 3 4 and 3 5 Tips To prevent your...

Страница 9: ...s adjusted to smaller size the contrast will be increased and the depth of field will be increased as well Turn up the intensity of the light if the image is too dim 3 8 Adjusting focus knob tension The tightness of the focus knob tension has been pre set at the factory If the mechanical stage drops by itself rotate the tension adjustment ring Fig 5 located inside the focus knob on the power switc...

Страница 10: ...0 0 17 spring 40X 0 65 160 0 17 spring Condenser Abbe NA 1 25 w iris aperture diaphragm and filter holder Rack and pinion focus adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides w stop Stage Double layer mechanical stage Dimension 5 1 2 x 4 5 16 140mm x 110mm Translation range 3 x 1 5 8 78mm x 40mm Illumination Transmitted 3W LED variable intensity Camera Built in USB...

Страница 11: ...17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Diopter adjustment is not set properly Readjust the diopter settings Condenser aperture is closed or open too much Open or close properly Condenser is positioned too low Position the condenser upward Spec...

Отзывы: