Caution (important):

 

  When you use the 100X objective to observe the specimen, you have to finish 

observing the specimen with the 4X, 10X, 40X objectives. 

  When you use the 100X objective to observe the specimen, you have to apply the 

immersion oil on the top of the slide cover. 

  When you apply the immersion oil with the 100X objective, do not let the immersion 

oil to contact with the dry objective lenses (especially the 40X). If the immersion oil is 
on the dry objectives lens, please use the camera cleaning kit to clean the objectives 
lens in time. The oil will damage the dry objective lenses. 

  After observing the specimen with the 100X objective, clean the 100X objective lens 

first. 

3.7  Adjusting condenser 

1)  Turn the condenser focus knob to raise or lower the condenser.   
2)  Raise the condenser when using high power objectives and lower it when using low 

power objectives. 

Note:   

  The centering of the condenser and the light axis of the objective are factory adjusted. 

Do not attempt to re-adjust.   

  The highest position of the condenser has been factory adjusted. Do not attempt to 

re-adjust. 

3.8  Adjusting iris aperture diaphragm 

Swing the iris diaphragm lever (

Fig. 9

) left or right to 

adjust the aperture size.   

Note:   

 

The iris diaphragm is designed to adjust the aperture 
size, not to adjust the brightness although the 
brightness will be changed when it's adjusted. When 
aperture is adjusted to smaller size, the contrast will be 
increased and the depth of field will be increased as 
well. Turn up the intensity of the light if the image is too 
dim

.   

 
 
 
 
 

Fig.8 

Iris Diaphragm Lever 

Fig.9 

Содержание M827L Series

Страница 1: ...User Manual Binocular Compound LED Microscope Model M827L Series MicroscopeNet com...

Страница 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 2 Components Illustration 3 3 Installation 4 4 Operation 6 5 Specifications 10 6 Optional Parts 11 7 Troubleshooting Guide 12...

Страница 3: ...of direct sunlight high temperature or humidity and dusty environments Ensure that the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto...

Страница 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Страница 5: ...Screw 19 Focus Stop Lever 4 Nosepiece 12 Microscope Body 20 Condenser Lock Thumb Knob 5 Objective 13 Focus Tension Ring 21 Condenser Focus Knob 6 Slide Holder 14 Coarse Focus Knob 22 Abbe Condenser 7...

Страница 6: ...mechanical stage is at its lowest position 2 Turn the caps counter clockwise to remove them from the nosepiece 3 Take the objectives out from the plastic cases and turn each one clock wise into the h...

Страница 7: ...take the eyepieces off and be certain that the head is securely locked by the thumb screw 2 6 Installing the mirror optional may not included in your package 1 Turn off the power switch and disconnec...

Страница 8: ...on the power switch 3 Rotate the brightness intensity dial to increase or decrease the brightness Caution A diffusion filter is attached beneath the condenser to get uniform light and protect your eye...

Страница 9: ...get a good focused image you may need to combine the focus knob adjustment and interpupillary distance adjustment along with eyepiece diopter adjustment stated in 3 3 and 3 4 5 You may now switch to...

Страница 10: ...s first 3 7 Adjusting condenser 1 Turn the condenser focus knob to raise or lower the condenser 2 Raise the condenser when using high power objectives and lower it when using low power objectives Note...

Страница 11: ...mera optional may not included in your package 1 Bring the microscope into focus by following the procedures in 3 5 2 Install the camera by following the procedures in 2 7 3 Open image observing sofew...

Страница 12: ...with iris diaphragm and filter holder Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with focus stop Minimum fine focusing adjustment at 0 002mm range...

Страница 13: ...0 x 480 0 3MP MS Windows Mac OS X 10 6 10 8 Included A3513U 1280 x 1024 1 3MP MS Windows Mac OS 10 8 and up Linux 2 6 or above A3520U 1600 x 1200 2 0MP A3530U 2048 x 1536 3 0MP A3550U 2592 x 1944 5 0M...

Страница 14: ...is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Diopter adjustment is not set pr...