Omax M824 Series Скачать руководство пользователя страница 6

 

 

www.microscopenet.com

 

 

- 4 - 

2.

 

Installation

 

 

2.1  Installation of the trinocular viewing head

 (

Fig.1

)

 

 

1)  Loosen the set screw on the top of the microscope 

body and remove the plastic cover. 

2)  Remove the cap on the dovetail of the trinocular 

viewing head. 

3)  Seat the dovetail of trinocular viewing head into the 

socket on the top of microscope body completely and 
tighten the set screw. 

Caution

:   

Do not release the viewing head from your hand grip until you are sure the viewing 
head is installed securely.

 

2.2  Installation of the photo tube 

1)  Remove the cap on the top of the viewing head. 
2)  Thread the photo tube into the top of the viewing head. 

2.3  Installation of the eyepieces   

1)  Remove the protective caps from the eyepiece tubes.   
2)  Insert the eyepieces into the eyepiece tubes. 

2.4  Installation of the objectives 

1)  Adjust the coarse focus knob until the mechanical stage is at its lowest position. 
2)  Turn the caps counter-clockwise to remove them from the nosepiece. 
3)  Take the objectives out from the plastic cases and turn each one clock-wise into the 

holes on the nosepiece. Install the 4X objective into the nosepiece first. Then in a 
counter-clockwise direction, rotate the nosepiece and install each succeeding higher 
magnification objective as shown in 

Fig. 2

 
 
 
 
 
 
 
 
 
 
 
 

Note:   

  Inspect the objectives frequently for dirt or oil; clean if necessary. 

  Use the 10X objective to initially focus the image of your specimen. 

  When changing the objective magnification, rotate the objective nosepiece until 

you hear a “click” sound or have a clear “in position” feeling. This ensures the 
objective is centered in the optical light path. 

Fig. 1 

Set Screw 

Fig.2 

 

Содержание M824 Series

Страница 1: ...User Manual Trinocular Compound Microscope Model M834 Series MicroscopeNet com...

Страница 2: ...1 Component Illustration 3 2 Installation 4 3 Operation 7 4 Specifications 10 5 Optional Parts 11 6 Troubleshooting Guide 13 7 Darkfield Condenser Installation and Operation Instructions 15 8 Phase C...

Страница 3: ...l electrical connectors power cord should be inserted into an electrical surge suppressor to prevent damage due to voltage fluctuations 6 Important the lamp lamphouse and adjacent parts will become ve...

Страница 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Страница 5: ...11 Microscope Base 19 Condenser Holder 4 Slide Holder 12 Eyepiece 20 X Y Stage Moving Knobs 5 Condenser Lock Thumb Screw 13 Diopter Adjusting Ring 21 Light Collector 6 Condenser Focus Knob 14 Eyepiec...

Страница 6: ...iece tubes 2 Insert the eyepieces into the eyepiece tubes 2 4 Installation of the objectives 1 Adjust the coarse focus knob until the mechanical stage is at its lowest position 2 Turn the caps counter...

Страница 7: ...ight housing 3 Take the eyepieces off and check if the view head is securely locked by the set screw Put the microscope down on its back gently 4 Open the cover on the bottom of the base by pull the k...

Страница 8: ...oto tube and then connect the camera to a computer via USB2 0 cable See Fig 6 3 The manual for the camera is on a CD or mini CD Refer to the manual to install the driver and software on to the compute...

Страница 9: ...il the specimen is close to the objective 2 Turn the coarse focus knob until the specimen is in focus 3 Use the fine focus knob to obtain a sharp image 4 To get a good focused image you may need to co...

Страница 10: ...sting condenser 1 Turn the condenser focus knob to raise or lower the condenser 2 The condenser is raised when using high magnification objectives and lowered when using low magnification objectives N...

Страница 11: ...microscope into focus by following the procedures in 3 3 2 Install the camera by following the procedures in 2 8 3 Open image observing software to examine more details see camera s manual 4 If the im...

Страница 12: ...Coaxial coarse and fine focusing knobs on both sides w stage upward stopper Minimum fine focusing adjustment at 0 002mm range 30mm Mechanical Stage Large double layer mechanical stage Dimension 5 1 2...

Страница 13: ...S Windows 32 64 bit Mac OS Included A1510 1280 x 1024 1 3MP A1520C 1600 x 1200 2 0MP A1530X 2048 x 1536 3 0MP A1550X 2592 x 1944 5 0MP A1590 3488 x 2616 9 0MP MS Windows 32 64 bit 2 Darkfield Condense...

Страница 14: ...Achromatic 20X with built in phase plate Achromatic 40X with built in phase plate spring Achromatic 100X with built in phase plate spring oil five positions 10 for 10X phase contrast objective 20 for...

Страница 15: ...17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Imm...

Страница 16: ...e condenser is not in the right position Lower or raise the condenser slightly to the position that the specimen is the brightest A191BOIL only Make sure the oil contact the bottom of the slide and th...

Страница 17: ...e condenser up or down slightly to get the best darkfield viewing Note The dry darkfield condenser is used with the dry objectives only The dry darkfield condenser works with the 4X 10X 40X objectives...

Страница 18: ...objectives a Take off all the objectives from the nosepiece b Install the phase contrast objectives onto the nosepiece following the steps in 2 4 2 Mounting the phase contrast condenser annular ring d...

Страница 19: ...on focusing etc following the instructions in this manual Note The phase contrast condenser will be working as a conventional Abbe condenser if the annular ring disk being put at B position 8 2 Phase...

Страница 20: ...he telescope The bright ring and dark ring should be coincided with each other as shown in Fig 21 d e If the ring images are not clear turn the top of telescope until both ring images are in focus f I...

Страница 21: ...is hard to find put a normal white print paper on the stage and under the objective then you will see the dark ring as shown in Fig 24 a h If the bright ring is still obscure as in Fig 24 b raise or l...

Отзывы: