USG-400, SONICBEAT & THUNDERBEAT 

Troubleshooting Guide

Error Code

Irregularity Description

Possible Causes

Remedial Actions

U511

SEAL short circuit
error

Output is generated while a
metallic object is touched.

1  Make sure that no metallic object is grasped.

Output in body fluids.

2 Withdraw excess fluids in the body.

Grasping section wear.

3     If the error window is still displayed, replace 

the THUNDERBEAT or SONICBEAT 
instrument.

U512

Open circuit error

The fully assembly instrument is
activated without holding tissue.

1 Re-grasp tissue and confirm activation.

Transducer and THUNDERBEAT 
instrument are not connected properly.

2  Check if the THUNDERBEAT instrument and 

transducer are connected. If not, connect 
them properly.

Transducer plug is not connected to the
THUNDERBEAT socket properly.

3  Confirm that the transducer plug is connected 

to the THUNDERBEAT socket on the front 
panel of the ultrasonic generator. If not, 
connect it properly.

Transducer malfunction.

4 Replace the transducer.

The THUNDERBEAT instrument
malfunction.

  5 Replace the THUNDERBEAT instrument.

Ultrasonic generator malfunction.

6 Contact Olympus.

U513

Time out error
(SEAL & CUT,
MAX, VAR)

Output time period is longer than
allowable by the system
(maximum output time: 60 s).

The maximum output time period has been
expired. Re-grasp the tissue and confirm
activation.

U514

Time out error
(SEAL incomplete)

The specified output time period has 
expired (maximum output time: 15 s).

Withdraw tissue or fluids from jaw and probe tip. 
Re-grasp tissue and confirm activation.

U601

Probe damage error 
(which is detected 
through PROBE CHECK)

The THUNDERBEAT or SONICBEAT 
instrument has not been replaced.

1   Replace the THUNDERBEAT or SONICBEAT 

instrument.

Transducer plug is not connected to the 
THUNDERBEAT or SONICBEAT socket 
properly.

2  Confirm that the transducer plug is connected 

to the THUNDERBEAT or SONICBEAT socket 
on the front panel of the ultrasonic generator. 
If not, connect it properly.

Transducer temperature is too high due 
to use immediately after autoclaving.

3 Stop output until the temperature cools down.

Transducer temperature is too high due 
to continuous output.

Transducer plug is not connected to the 
THUNDERBEAT or SONICBEAT socket 
properly.

4 Replace the transducer.

Transducer plug is not connected to the 
THUNDERBEAT or SONICBEAT socket 
properly.

5  Replace the THUNDERBEAT or  

 SONICBEAT  instrument.

Transducer plug is not connected to the 
THUNDERBEAT or SONICBEAT socket 
properly.

6 Contact Olympus.