Troubleshooting Guide

Replace instrument: USG-400

 

IFU page 128-129

1

 

Without turning off the ultrasonic generator or disconnecting the transducer plug from the transducer socket of the  

 

ultrasonic generator, disconnect the handle from the transducer and replace it with a new THUNDERBEAT or  

 

SONICBEAT handle.

2

 

Probe check push button 

 

Press the “PROBE CHECK” push button on the right side of the front panel. 

 

The confirmation window (Probe Check) is displayed.

Executing the probe check 

 

With the grasping section open, press and hold the handswitch or footswitch pedal for 

 

three seconds and then release it.

 

  • The probe check starts.

 

• The information window (Checking) is displayed.

 

•  When the check is completed, the information window (Checking) disappears and the probe check  

result is displayed.

Error window 

(U504: Probe damage error)

Confirmation window 

(Probe Check)

)