UES-40
T
roub
le
sh
oot
ing
2
-6
IS
SUE4
(From previous
page)
Error code dis-
played
High-frequency current monitor dam-
age (Cut 01)
—
1
3
2
—
—
Monitor board (J7) – Relay board (J2)
—
Used where the high-frequency cur-
rent is prone to leakage. (Touching
metal parts of patient’s bed or objects
such as external devices that have
fallen to the ground; used simulta-
neously with high-frequency grounded
electric scalpels [semi-floating], etc.)
Monitor board (J16) – Control board (J1)
High-frequency voltage-monitor dam-
age (Cut 02)
2
1
4
3
—
5
Converter – Oscillation board (J1)
—
—
Converter – Oscillation board (J5)
Monitor board (J3) – Oscillation board (J7)
Monitor board (J6) – Oscillation board (J3)
Oscillation board (J4) – Control board (J5)
Monitor board (J7) – Relay board (J2)
Monitor board (J16) – Control board (J1)
Waveform-circuit damage (Cut 04)
—
—
—
1
—
—
—
—
—
Converter (Power-supply) 1 damage
(Cut 08)
3
—
—
1
—
2
Oscillation board (J4) – Control board (J5)
—
—
Converter (Power-supply) 2 damage
(Cut 10)
3
—
—
1
—
2
Oscillation board (J5) – Converter
—
—
Oscillation board (J4) – Control board (J5)
Converter (Power-supply) 3 damage
(Cut 20)
3
—
—
1
—
2
Oscillation board (J5) – Converter
—
—
Oscillation board (J4) – Control board (J5)
Patient-plate-circuit damage (Cut 40)
—
1
3
2
—
—
Monitor board (J5) – Relay board (J1)
—
—
Monitor board (J16) – Control board (J1)
Problem
Parts That Might Have Caused the Problem
Oscillation
board
Monitor
board
Relay
board
Control
board
Panel
sheet
Converter
Harnesses
Other
Other than main unit
*
1
Содержание UES-40
Страница 1: ...ISSUE4 2008 01 MAINTENANCE MANUAL UES 40...
Страница 2: ......
Страница 6: ......
Страница 28: ......
Страница 36: ......
Страница 53: ...ISSUE4 4 1 Inspection Chapter 4 Inspection Chapter 4 Inspection 1 Preparation 4 2 2 Inspection Procedure 4 2...
Страница 55: ...ISSUE4 5 1 Parts List Chapter 5 Parts List Chapter 5 Parts List 1 Exploded Parts Diagram 5 2 2 Parts List 5 6...
Страница 62: ......
Страница 63: ......