UES-40
ISSUE4
1-5
Product Specification
2-1-8
Output terminals
• Bipolar system
Bipolar cord port
2 ports
• Monopolar system
Active port
1 port
Handpiece port
2 ports
• Saline system
Dedicated cord port
1 port
Different control systems are available, depending on the output terminal. The output termi-
nals and their corresponding controls are listed below.
2-1-9
Recommended timing
Output (ON) timing: within 10 seconds; output stop (OFF): 30 seconds or more
2-2
Safety functions
2-2-1
Patient-plate monitor (for monopolar systems only)
Detects connection problems or disconnection of the single type patient plate, as well as
loosening of split type patient plates. If problems are detected, the following operations are
performed.
• Warning indicator is lit.
• Error code is displayed on output indicator.
• Audible warning is issued.
• High-frequency output is stopped.
System
Output terminal
Output
option
Control system
Remarks
Bipolar
Bipolar 1 port
Coag, Cut
Foot switch for UES-40
(MAJ-1258)
Bipolar output can be
selected using the Foot-
switch select switch.
Bipolar 2 port
Coag only
Bipolar foot switch for
UES-40 (MAJ-1259)
Saline
Saline port
Coag, Cut
MAJ-1258
The saline port detects the
saline cable to be connected
and the output terminal is
automatically selected.
Monopolar
Handpiece 1 port
Coag, Cut
MAJ-1258
handpiece
Monopolar output can be
selected using the Foot-
switch select switch (except
during control using the
handpiece)
Coag, Cut
MAJ-1258
Handpiece 2 port
Coag, Cut
Handpiece
Active port
Содержание UES-40
Страница 1: ...ISSUE4 2008 01 MAINTENANCE MANUAL UES 40...
Страница 2: ......
Страница 6: ......
Страница 28: ......
Страница 36: ......
Страница 53: ...ISSUE4 4 1 Inspection Chapter 4 Inspection Chapter 4 Inspection 1 Preparation 4 2 2 Inspection Procedure 4 2...
Страница 55: ...ISSUE4 5 1 Parts List Chapter 5 Parts List Chapter 5 Parts List 1 Exploded Parts Diagram 5 2 2 Parts List 5 6...
Страница 62: ......
Страница 63: ......