12
Rotate the light intensity adjustment dial E to adjust the light
intensity.
· Rotating the light intensity adjustment dial E allows you to
change the numerical characters (Min: 010, Max: 100) on the
counter D in increments of 5.
· The larger the numerical characters on the counter D
become, the brighter the lamp becomes.
When using U-LGPS control software:
· You cannot adjust the light intensity using the light intensity
adjustment dial E . Adjust the light intensity on the U-LGPS
control software.
1
TIP
3
Adjusting the brightness
E
D
The relation between the color of the pilot lamp B and the display
on the counter D are listed as follows.
4
Pilot lamp color and counter display
D
B
Pilot lamp B
Display on counter D
Color
Status
Display
Description
Green Lamp is OFF
010-100
Brightness of the
lamp when the
lamp is ON
Blue Lamp is ON
Red
Error in U-LGPS (Lamp
turns OFF automatically)
c01-c23 *
Details of the
error in U-LGPS
When using U-LGPS control software:
The color of the pilot lamp B changes as the table shown above,
and the display of the counter D always shows 000.
* The procedure to recover from the error is described below.
Display
Recovery procedure
Page
c06
Insert the liquid light guide into the liquid
light guide insertion port of U-LGPS properly.
15
Other than
c06
Set the power switch A to
(OFF), wait
for a while and turn ON the lamp.
11
If you cannot recover from the error according to the above
procedure, contact Olympus for assistance.
Содержание U-LGPS