25
SZX7
Place a specimen on the stage plate, hold the cup stage by the edge,
and tilt the cup stage slowly.
If the specimen slips on the stage plate, hold the specimen
with the provided specimen holder.
1
TIP
To fix a container such as a petri dish, insert the provided tube into
the specimen holder to fix the container by pinching.
3
Operation
·
Do not touch the sliding surface on the cup stage and seat with a hand. If the sliding surface are contaminated
with oil, etc., wash with a neutral detergent before use.
·
If an eccentric load of more than 20 grams is applied to the edge of the cup stage, it may move spontaneously.
·
When a tall specimen is placed and the cup stage is tilted, the specimen may go out of focus. In this case,
adjust the focus again.
NOTE