background image

13

SZX10

TROUBLESHOOTING GUIDE

Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a problem
occurs, please review the following list and take remedial action as needed. If you cannot solve the problem after
checking the entire list, contact Olympus.

9

Problem

Cause

Remedy

Page

1. The left and right fields of view

do not coincide.

The interpupillary distance is adjusted
improperly.

Adjust it correctly.

The diopter is adjusted incorrectly.

Adjust it correctly.

The left and right eyepieces are different.

Use a pair of matched eyepieces.

2. The field of view is obscured

partially or illuminated unevenly.

The aperture iris diaphragm is stopped
down excessively.

Open the aperture iris diaphragm.

The trinocular observation tube and/or in-
termediate attachments are installed im-
properly.

Mount them properly.

The light path selector knob is in an inter-
mediate position.

Set it correctly to the desired position.

3. The monitor image is cut off

partially.

The light path selector knob is pulled out
incompletely.

Pull it out all the way.

4. Dirt or dust is visible in the field

of view.

Dirt/dust on the specimen.

Clean thoroughly.

Dirt/dust on eyepiece.

5. Details of observed image are

solid.

Clean thoroughly.

The aperture iris diaphragm is stopped
down excessively.

Open the aperture to a proper diameter.

6. Visibility of observed image is

poor.
· Image is not sharp.
· Contrast is insufficient.

The objective is tilted.

Screw it correctly until it stops.

The objective is dirty.

Clean thoroughly.

The top and/or bottom lenses of the mi-
croscope body are dirty.

The bottom lens of the observation tube
is dirty.

7. Zooming causes defocusing

of the observed image.

The eyepiece diopter is adjusted improperly.

Adjust it correctly.

The focus adjustment is in accurate.

Adjust focus at a high magnification.

8. The coarse focus adjustment

knob does not rotate smoothly.

The rotation tension of the knob is set too
high.

Decrease the rotation tension to an opti-
mum level.

9. The microscope body drops

spontaneously, causing the
focusing to be deviated dur-
ing observation.

The rotation tension of the coarse focus
adjustment knob is set too low.

Increase the rotation tension to an opti-
mum level.

The microscope body dropped spontaneously
because the weight exceeded 10 kg.

Use a focusing assembly withstanding
heavy loads. (SZX2-FOF: 2.7 to 15 kg. SZX2-
FOFH: 8 to 25 kg.)

9

20

8

19

11

11

3

3

8

9

9

7

7

14

19

3

Содержание SZX10

Страница 1: ...mance and to familiarize yourself full with the use of this microscope we recommend that you study this manual thoroughly before operating the microscope Retain this instruction manual in an easily ac...

Страница 2: ...This device complies with the requirements of directive 98 79 EC concerning in vitro diagnostic medical devices CE marking means the conformity to the directive...

Страница 3: ...the Rotation Tension of the Coarse Focus Adjustment Knob 2 Engaging and Disengaging the Zooming Knob Click Stop Function 3 Adjusting the Aperture Iris Diaphragm 4 Zoom Magnification Indication 1 Adjus...

Страница 4: ...t Adjuster SZX EPA 21 9 2 Drop Prevention Collar SZX R and Auxiliary Pillar SZH P400 SZH P600 22 9 3 Revolving Nosepiece SZX2 2RE10 23 25 9 4 BX Stage Adapter Type 1 SZX STAD1 26 28 9 5 BX Stage Adapt...

Страница 5: ...not turn over 2 To prevent the microscope body from turning over its left right pivot angle must be limited to 30 as shown in Fig 1 The tilt of the desktop surface should be no more than 5 3 Sufficien...

Страница 6: ...it hits a limiting mechanism the internal mechanism may be damaged 2 If the knobs on the left and right are ro tated in opposite directions the internal mechanism will be damaged The ro tation tension...

Страница 7: ...ions or reduced performance 4 When not using the microscope keep it covered with the dust cover provided If a module containing a source of heat is mounted on the microscope wait until it has cooled d...

Страница 8: ...X BI30 BI45 Trinocular observation tube SZX2 TR30 TR30PT Tilting binocular tube SZX TBI Tilting trinocular observation tube SZX2 TTR TTRPT Focusing assembly Coarse fine focusing assembly SZX2 FOF Focu...

Страница 9: ...F screw P 8 Engages or disengages the click stop function for each zoom mag nification Light path selector knob P 11 Trinocular tube only Zooming knob 0 63X to 6 3X Coarse focus adjustment knob rotati...

Страница 10: ...ring the microscope into focus by rotating the coarse focus adjustment knob 5 Rotate the zooming knob to the desired magnification and precisely focus the microscope on the specimen with the coarse fo...

Страница 11: ...heavy load SZX2 FOFH The rotation tension of the coarse focus adjustment knob can be adjusted with the rotation tension adjustment ring Do not rotate the knobs on the left and right in opposite direc...

Страница 12: ...d image as well as the depth of focus However setting the aperture iris diaphragm too narrowly degrades reso lution 1 Adjust the aperture iris diaphragm ring to the left or right Rotating the ring tow...

Страница 13: ...nd looking only into the right eyepiece focus the specimen using the coarse and fine focus adjustment knobs 5 Rotate the zooming knob to the lowest magnification look only into the left eyepiece and f...

Страница 14: ...sure to perform magnifi cation compensation when using this eyepiece in measurement When the micrometer disk is engaged in their light path the light path length will extend and the diopter scale may...

Страница 15: ...o a stop position Do not attempt to force the knob past the stop position Applying excessive force could destroy the mechanism Binocular 50 TV photo 50 Binocular 50 SZX2 TR30PT SZX2 TTRPT Right Binocu...

Страница 16: ...bservation photomicrography areas when the WHSZ10X eyepieces with FN 22 are used 1 in CCD 2 3 in CCD 1 2 in CCD WHSZ10X FN 22 When the 0 5X TV adapter is used When the 1X TV adapter is used 2 Mounting...

Страница 17: ...in the field of view Dirt dust on the specimen Clean thoroughly Dirt dust on eyepiece 5 Details of observed image are solid Clean thoroughly The aperture iris diaphragm is stopped down excessively Op...

Страница 18: ...PT Coarse focus adjustment knob only Coarse focus adjustment knob stroke 80 mm Coarse knob stroke per turn 36 8 mm Load 8 to 25 kg Load 2 7 to 15 kg Max load 10 kg SZX BI30 BI45 SZX TBI Binocular obse...

Страница 19: ...crometer disk can be inserted Note The micrometer area outside the FN is invisible WHSZ10X H FN 22 diopter adjustment ring provided WHSZ15X H FN 16 diopter adjustment ring provided WHSZ20X H FN 12 5 d...

Страница 20: ...on magnification Objective magnification x Zoom magnification x Eyepiece magnification Observation area Eyepiece FN Objective magnification x Zoom magnification mm 4 73X 46 6 7 09X 33 9 9 45X 26 5 14...

Страница 21: ...provided with the zoom microscope body The Allen screwdriver can be stored in the pil lar Do not use the Allen wrench for a purpose other than installing the pillar and the zoom microscope body Eyepi...

Страница 22: ...w Fig 16 Insert vertically and slowly Do not apply excessive force 2 Lower the focusing assembly until it stops then tighten the focusing assembly clamping knob Fig 16 To prevent the microscope from t...

Страница 23: ...objective tip Place the cap of the objective mount thread or a notebook etc on the stand to absorb the shock even when the objective drops While holding the tip of the objective mount it on the objec...

Страница 24: ...at the edge nearest to the pillar with your fingertip The opposite end will rise from the base so the stage plate can be picked up easily 8 Mounting the Specimen Holder Fig 23 Use the specimen holder...

Страница 25: ...o other intermediate attachment is used Observation tube clamping screw 2 Assembly Fig 25 1 Using the Allen screwdriver provided with the SZX2 microscope body remove the observation tube 2 Mount the e...

Страница 26: ...ort work ing distance is used the drop prevention collar does not have to be mounted if you move the microscope body downward However this reduces safety that is assured by using the SZX R drop preven...

Страница 27: ...ve mounts Clamping screws M3 6 mm long x 3 1 Remove the observation tube from the zoom microscope body 2 Remove the zoom microscope body and focusing assembly from the pillar and place them upside dow...

Страница 28: ...d install the observation tube in the original position 6 Attach two objectives onto their respective objective mounts by screw ing Fig 28 Hold the objective and gently rotate it until a click positio...

Страница 29: ...light path can be used as the photo light path of optical axis aligned photomicrography 1 Rotate the desired objective counterclockwise 15 till the click posi tion 2 Return the microscope frame to th...

Страница 30: ...e Applicable Objectives Restrictions Standard base SZX2 ST SZX ST 0 5X to 2X None Transmitted illuminator stand SZX2 ILLK SZX ILLK High class illuminator stand SZX2 ILLB SZX ILLB2 Brightfield darkfiel...

Страница 31: ...ting threaded holes x 2 Base Mounting the Polarizer SZX PO When simplified transmitted polarized light observation is required install the SZX PO polarizer on the SZX STAD1 BX stage adapter type 1 To...

Страница 32: ...ping knobs fit the analyzer into the objective tip and tighten the clamping knobs by positioning them on both sides 2 Without a specimen rotate the analyzer rotation ring to make the field of view dar...

Страница 33: ...ether with the drop prevention collar The U SVL or U SVR BS stage for the BX series microscopes can also be mounted but the operability deteriorates The U SVLB and U SVRB cannot be used due to their l...

Страница 34: ...e illuminator remove the frosted filter and while observing through the eyepiece tilt the mirror gradually until optimum contrast can be obtained Do not project the image of the external light source...

Страница 35: ......

Страница 36: ...ny 3500 Corporate Parkway P O Box 610 Center Valley PA 18034 0610 U S A One Corporate Drive Orangeburg NY 10962 U S A 491B River Valley Road 12 01 04 Valley Point Office Tower Singapore 248373 31 Gilb...

Отзывы: