13
SZX-SDO2
1
Mounting the pillar
2
Mounting the focusing assembly
When the primary observer is to sit on the right side, the pillar
support should be moved to the left side.
Using the Allen wrench (3 mm across flats) provided with the base,
fully loosen the 2 pillar support clamping screws a .
Hold the pillar b with the hex screwdriver storage hole made of white
rubber up, and gently insert the pillar all the way into the mounting
hole until it stops.
Using the Allen wrench, tighten the 2 clamping screws a securely.
TIP
1
2
3
Fully loosen the focusing assembly clamping knob a . While holding
the focusing assembly with both hands, insert the pillar c into the
mounting hole b .
Insert gently without applying excessive force.
After inserting the focusing assembly until it reaches the stop position,
secure it with the focusing assembly clamping knob a .
·
To prevent the microscope body from turning over, be sure
to mount the focusing assembly so that it is located on the
front as shown by “ ” in figure below and clamp securely.
The microscope will turn over if the focusing assembly is
mounted facing the rear.
30° or less
·
When attaching the focusing assembly on the top of the
pillar of the SZX2-STL2 base, be sure to insert the SZX-R
drop prevention ring (optionally available) immediately
below the focusing assembly without leaving any clearance.
1
NOTE
2
CAUTION
5 -2 Detailed assembly procedure
a
b
a
b
c
CORRECT
INCORRECT