11
SZX-SDO2
Large base SZX2-STL2
Item
Specification
1. Base
Size
500 x 350 mm
Pillar locations
2
2. Pillar
Height
400 mm (from base top surface)
External diameter
32 mm dia., f 7
-0.020
-0.041
3. Installation of stage adapter
Clamping onto base top surface using screws.
Clamped at 2 locations (pillar mounting locations)
4. Dimensions
500 (W) x 350 (D) x 424 (H) mm
5. Weight
9.5 kg (20.9 lb)
Operating environmen
· Indoor use.
· Altitude: Max 2,000 m.
· Ambient temperature: 5°C to 40°C. (41°F to 104°F)
· Maximum relative humidity 80% for temperatures up to 31°C (88°F), decreasing linearly through 70% (at 34°C/93°F),
60% (at 37°C/99°F) to 50% (at 40°C/104°F).
· Supply voltage fluctuation: ±10%.
· Pollution degree: 2 (in accordance with IEC60664).
· Installation/Overvoltage category: II (in accordance with IEC60664).