3
Important safeguards/notices for use in the
medical environments
1.
All the equipments connected to this unit shall be
certified according to Standard IEC601-1, IEC950,
IEC65 or other IEC/ISO Standards applicable to the
equipments.
2.
When this unit is used together with other equipment in
the patient area*, the equipment shall be either powered
by an isolation transformer or connected via an
additional protective earth terminal to system ground
unless it is certified according to Standard IEC601-1
and IEC601-1-1.
* Patient Area
3.
The leakage current could increase when connected to
other equipment.
4.
The operator should take precautions to avoid touching
the rear panel input and output circuitry and the patient
at the same time.
5.
Model OEV143/OEV203 is a video monitor intended
for use in a medical environment to display video
pictures from EVIS video system center or other video
system.
Symbols on the unit
Symbol
This symbol indicates
Main power switch.
Press to turn the monitor
on or off.
The equipotential terminal
which brings the various
parts of a system to the
same potential.
Functional earth terminal
Alternating current
Attention, consult
ACCOMPANYING
DOCUMENTS
R
1.5m
Warning on power connection
Use a proper power cord for your local power supply.
Location
Front panel
Rear panel
Rear panel
Rear panel
Rear panel
English
EN
United State
Canada
Continental Europe
Japan
Plug type
HOSPITAL GRADE*
LP-34A or LP-34
VM1050
Female end
Min.Type SJT
LS-60
VM1010
Cord type
Min.18 AWG
H05VV-F
PVCTF
Minimum cord set rating
10A/125V
10A/125V
10A/250V
12A/125V
Safety approval
UL Listed
CSA
VDE
DENAN-HO
*Note: Grounding reliability can only be achieved when the equipment is connected to
an equivalent receptacle marked ‘Hospital Only’ or ‘Hospital Grade’.