Figure 5
‑
213 The scan of the two grooves
3.
Press the ANGLE key (
), and then rotate the signal until the OD grooves are
split by the vertical cross-hair (see Figure 5-214 on page 274).
Figure 5
‑
214 Adjusting the signal phase
4.
Press the GAIN key (
), and then decrease the gain until the 60 % wall-loss
flaw signal reaches approximately 90 % of vertical height.