Olympus MU-1 Скачать руководство пользователя страница 8

Product Outline

1-2

ISSUE6

MU-1

3. Restrictions

(1)

OES scopes and EVIS scopes can be used.

(2)

Unless the AW channel cleaning adapter, leak tester and chemical tank are used, draining
water from the air and water channel, leak inspections and feeding rubbing alcohol for
drying the OES scope and EVIS scope are not possible.

(3)

In order to drain the water feeding channel, cover the water supply connector with a finger.
(Same as with the former model.)

(4)

The ambient operation conditions are as follows:
a) Temperature:  10 to 40 ºC
b) Humidity:  30 to 85%
c) Atmospheric pressure:  700 to 1060 hPa
d) Not to be used in an atmosphere with inflammable gases.

(5)

The power supply conditions are as follows:

a) Rated power supply voltage:  100 V, 120 V, 220 to 240 V (according to the intended

market)

b) Rated input current:  0.4 A, 0.4 A, 0.2 A
c) Frequency:  50/60 Hz
d) Allowed voltage fluctuation: ±10%

Содержание MU-1

Страница 1: ...ISSUE6 2016 04 149P 0015 MU 1 MAINTENANCE UNIT Maintenance Manual...

Страница 2: ......

Страница 3: ...dified product The information contained in this manual is subject to change at any time without notice Later editions of this manual will contain revised and updated material Applicable Unit MAINTENA...

Страница 4: ......

Страница 5: ...ous Components 1 5 6 System Chart 1 7 Chapter 2 Specifications 2 1 Chapter 3 Installation and Connection 3 1 Chapter 4 Care Storage and Disposal 4 1 Chapter 5 Inspection 5 1 1 Inspection Check Sheet 5...

Страница 6: ......

Страница 7: ...r an air bubbles 3 This product can be combined with the AW channel cleaning adapter and chemical tank to feed rubbing alcohol into the OES scope and EVIS scope air and water channel in order to dry t...

Страница 8: ...o drain the water feeding channel cover the water supply connector with a finger Same as with the former model 4 The ambient operation conditions are as follows a Temperature 10 to 40 C b Humidity 30...

Страница 9: ...temperature of 20 to 25 C 4 1 4 Control Method By turning the power switch ON OFF 4 1 5 Hook Installation Can be hung down by connecting the hanger to the hook installed in a wall The hook is screwed...

Страница 10: ...Product Outline 1 4 ISSUE6 MU 1 4 2 3 Atmospheric Pressure 700 to 1060 Pa 4 2 4 Inflammable Atmosphere Use in an inflammable atmosphere is prohibited...

Страница 11: ...air is supplied through the connecter Power supply switch Pushing the top will turn on the power supply and the power supply switch lamp will light at the same time The pump will start to operate and...

Страница 12: ...Product Outline 1 6 ISSUE6 MU 1 5 2 Back Power supply inlet Breaker...

Страница 13: ...1 6 System Chart Water resistant Cap MaintenanceUnit MU 1 Alcohol Container MB 264 only EVIS Videoscope 100 200 series OES Fiberscope 10 20 30 series Leakage Tester MB 155 OES Fiberscope EVIS Videosco...

Страница 14: ......

Страница 15: ...ISSUE6 2 1 Specifications MU 1 Chapter 2 Specifications Refer to the Specifications of Instruction Manual...

Страница 16: ......

Страница 17: ...ISSUE6 3 1 Installation and Connection MU 1 Chapter 3 Installation and Connection Refer to the Installation and Connection of Instruction Manual...

Страница 18: ......

Страница 19: ...ISSUE6 4 1 Care Storage and Disposal MU 1 Chapter 4 Care Storage and Disposal Refer to the Care Storage and Disposal of Instruction Manual...

Страница 20: ......

Страница 21: ...eck item may change with the destination and specifications There may be a check item which is on this list and is not in the instruction manual Please refer to the instruction manual Chapter of Inspe...

Страница 22: ...MU 1 MU 1 INSPECTION CHECK SHEET Serial number Year Month Day 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Inspecting the Operation Inspector Day 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Inspecting the...

Страница 23: ...esting instrument for the measurement methods 2 Grounding Leak Current Criteria Grounding leak currents should not exceed the values indicated below Inspection procedure Measure the leak current of th...

Страница 24: ...and grounding line should be 0 2 ohm or less Inspection procedure Measure the resistance between the grounding terminal in the removable power supply cord and the screw that is farthest from the inlet...

Страница 25: ...pected by Serial number Checked by Item Criteria Result 2 1 1 External leak current Normal Condition 100 A or less Single Fault 500 A or less 2 1 2 Grounding leak current Normal Condition 500 A or les...

Страница 26: ......

Страница 27: ...ISSUE6 7 1 Troubleshooting MU 1 Chapter 7 Troubleshooting Refer to the Troubleshooting of Instruction Manual...

Страница 28: ...2951 Ishikawa machi Hachioji shi Tokyo Japan Medical Service Engineering Department Medical Repair Service Division PRINTED IN JAPAN FIRST EDITION 10 2003...

Отзывы: