ISSUE6
6-3
Safety Check
MU-1
4. Safety Check Sheet
Electrical Safety Inspection
Date of inspection
Product
MU-1
Inspected by
Serial number
Checked by
Item
Criteria
Result
2-1-1
External leak current
Normal Condition
100
A or less
Single Fault
500
A or less
2-1-2
Grounding leak current
Normal Condition
500
A or less
Single Fault
1000
A or less
2-1-3
Grounding Resistance
Between the external case and grounding line.
0.2
Ω
or less