background image

Chapter 2  Instrument Nomenclature

7

MAINTENANCE UNIT MU-1

Chapter 2 Instrument Nomenclature

2.1

Instrument Nomenclature

Front Panel and Side Panel

Hook

Used to attach the Maintenance 
Unit to a wall. Use the screws to 
attach the Hook to a wall near a 
sink.

Output Socket

Accepts the endoscope or 
leakage tester and feeds air.

Screw ×2

Power Switch

Switches the power ON and OFF. 
The power switch lamp will 
illuminate and the unit emits air 
from the air pipe in the output 
socket when the power is on.

Ventilation Grills

Feet

Attaches the Maintenance Unit to 
the hooks or attaches the MB-264 
Alcohol Container to the 
Maintenance Unit. Brackets are 
provided on both sides of the unit.

Attachment Brackets 

(provided on both sides)

Содержание MU-1

Страница 1: ...USA CAUTION Federal law restricts this device to sale by or on the order of a physician INSTRUCTIONS MAINTENANCE UNIT MU 1...

Страница 2: ......

Страница 3: ...hapter 3 Preparation and Inspection 9 3 1 Preparation of the Equipment 9 3 2 Preparation and Inspection of the Maintenance Unit 11 3 3 Preparation and Inspection of the Endoscope and Disinfectant Cont...

Страница 4: ...Contents ii MAINTENANCE UNIT MU 1 Specifications 26...

Страница 5: ...Safety related labels are attached to the instrument at the locations shown below If labels are missing or illegible contact Olympus Rear side Electrical Rating Plate Product name rated voltage and se...

Страница 6: ...2 Labels and Symbols MAINTENANCE UNIT MU 1 Back cover of this instruction manual Manufacturer Authorized representative in the European Community...

Страница 7: ...l information on using this instrument safely and effectively Before use thoroughly review this manual and the manuals of all equipment which will be used during the procedure and use the instruments...

Страница 8: ...damage can result If the problems appears minor refer to Chapter 6 Troubleshooting If the problem cannot be resolved by using Chapter 6 contact Olympus Signal Words The following signal words are use...

Страница 9: ...inspect or use this instrument with wet hands Never use the Maintenance Unit in a flammable or high concentration oxygen environment and keep all flammable gases and liquids or high concentration oxy...

Страница 10: ...king the Package Contents Match all items in the package with the components shown below Inspect each item for damage If the instrument is damaged a component is missing or you have any questions do n...

Страница 11: ...ut Socket Accepts the endoscope or leakage tester and feeds air Screw 2 Power Switch Switches the power ON and OFF The power switch lamp will illuminate and the unit emits air from the air pipe in the...

Страница 12: ...nstrument Nomenclature MAINTENANCE UNIT MU 1 Rear Panel Symbols and Descriptions Power Switch ON OFF Power Cord Receptacle Circuit Breakers Protects the Maintenance Unit and other equipment from elect...

Страница 13: ...ng to the System Chart in the Appendix prepare this instrument other equipment that are used with this instrument paper towels basin gauze and protective equipment such as eye wear a face mask moistur...

Страница 14: ...ainer Maintenance Unit MU 1 Endoscope Leakage Tester MB 155 Optional Water resistant Cap provided with the endoscope EVIS Videoscope only Personal Protective Equipment Maintenance Unit MU 1 Endoscope...

Страница 15: ...to the same breaker Do not place the Maintenance Unit on its side or upside down When installing the Maintenance Unit on the hooks confirm that the screws which attach the hooks to the wall are not lo...

Страница 16: ...h the Hook to the wall are not loose and that the Hook does not move when the unit is attached to it see Figure 3 1 Figure 3 1 Connecting the Power Supply The plug of the power cord that follows the e...

Страница 17: ...igure 3 2 Inspecting the Operation 1 Turn the Power Switch ON 2 Confirm that the Power Switch lamp lights and that the pump starts operating 3 Confirm that air comes out of the lower hole in the Outpu...

Страница 18: ...once If the power is still not supplied to the Maintenance Unit after resetting the Circuit Breakers immediately turn the Power Switch OFF disconnect the Power Cord from a Wall Mains Outlet and contac...

Страница 19: ...in a solution Leakage Testing Refer to the instruction manuals of the Leakage Tester and endoscope for directions on Leakage Testing Remove water from the Air Water Channel For endoscope with Air Wate...

Страница 20: ...es only when irrigating the channel with 70 ethyl or isopropyl alcohol Operate to the Alcohol Container MB 264 according to the directions given in it s instruction manual 4 Fill the Alcohol Container...

Страница 21: ...hock Use personal protective equipment to guard against dangerous chemicals and infectious material During operation wear appropriate protective equipment such as eye wear a face mask moisture resista...

Страница 22: ...ethyl or isopropyl alcohol facilitates drying inside the Air Water channel When Irrigating With 70 Ethyl or Isopropyl Alcohol 1 Turn the Power Switch ON 2 Depress the button on the AW Channel Cleanin...

Страница 23: ...with your finger Press the AW Channel Cleaning Adapter button and release it Repeat this procedure at intervals of approximately 30 seconds until no more water or 70 ethyl or isopropyl alcohol exit t...

Страница 24: ...pe the outside surface with hard or abrasive wiping material The surface will be scratched If the Maintenance Unit becomes soiled perform the following cleaning procedure immediately after use Stains...

Страница 25: ...ctromagnetic radiation near microwave medical treatment equipment short wave medical treatment equipment MRI equipment or radio Damage to the Maintenance Unit may result Store the equipment in the hor...

Страница 26: ...ts If an accessory part becomes damaged contact Olympus to purchase a replacement 6 1 Troubleshooting Guide Irregularity Description Possible Cause Solution The pump does not work The Power Cord is no...

Страница 27: ...nnel When returning the Maintenance Unit for repair include a description of the Maintenance Unit malfunction or damage and the name and telephone number of the individual at your location who is most...

Страница 28: ...this instrument are listed below New products released after the introduction of this instrument may also be compatible for use in combination with this instrument For further details contact Olympus...

Страница 29: ...UNIT MU 1 Water resistant Cap EVIS Videoscope AW Channel Cleaning Adapter Maintenance Unit MU 1 Alcohol Container MB 264 only EVIS Videoscope 100 200 series OES Fiberscope 10 20 30 series OES Fibersco...

Страница 30: ...ive Humidity 30 85 without condensation Air Pressure 700 1060 hPa 0 7 1 1 kgf cm2 10 2 15 4 psia Item Specification Power Supply Voltage 120 V AC Voltage Fluctuation Within 10 Frequency 50 60 Hz Frequ...

Страница 31: ...The last digit of the year of manufacture is the second digit of the serial number EMC Applied standard IEC 60601 1 2 2001 This instrument complies with the standards listed in the left column CISPR 1...

Страница 32: ......

Страница 33: ......

Страница 34: ......

Страница 35: ...reproduced or distributed without the express written permission of OLYMPUS MEDICAL SYSTEMS CORP OLYMPUS is a registered trademark of OLYMPUS CORPORATION Trademarks product names logos or trade names...

Страница 36: ...Valley Road 12 01 04 Valley Point Office Tower Singapore 248373 Fax 6834 2438 Telephone 6834 0010 A8F Ping An International Financial Center No 1 3 Xinyuan South Road Chaoyang District Beijing 100027...

Отзывы: