For vacuum generator suction cups (optional), connect and turn on the air
supply to the vacuum generator, and check that there is sufficient vacuum to
hold both cups securely on the inspection surface. Be sure to use a reliable,
uninterrupted supply of air with sufficient flow and pressure to continuously
secure the cups on the surface during inspection (see Table 4 on page 54 for
specifications).
Figure 2
‑
38 Attaching manual suction cups to the inspection surface
4.
Set the Y-axis level:
a)
Loosen the thumbscrew used to lock the Y-axis support.
b)
Adjust the level of the Y-axis.
c)
When the Y-axis is level, tighten the thumbscrew (see Figure 2-39 on page 43).