DMTA029-01EN, Rev. E, December 2017
Scanner Assembly and Setup
33
2.2.5
Replacing a Conventional Ultrasonic Transducer
To replace a conventional ultrasonic transducer, the transducer-holder assembly must
be removed and partially disassembled.
To replace a conventional ultrasonic transducer
1.
Turn off the couplant flow.
2.
Disconnect the couplant tubes from the couplant source, and from the wedge, if
necessary.
3.
Disconnect the transducer cable.
4.
Remove the scanner from the inspection surface.
5.
Using a flat screwdriver, at the back of the bearing arm, unscrew the screw to free
the yoke from the bearing arm (see Figure 2-26 on page 33).
Figure 2
‑
26 The screw that holds the yoke on the bearing arm
6.
Remove the yoke.
7.
Using a hexagonal key, remove the screw that holds the yoke parts together (see
Figure 2-27 on page 34).