DMTA-10083-01EN, Rev. 3, June 2017
Appendix A
46
Table 7
Pulser
Parameter
Specifications
Pulser
Tunable square wave
PRF
10 Hz to 2000 Hz in 10 Hz increments
Energy settings
100 V, 200 V, 300 V, or 400 V
Pulse width
Adjustable from 25 ns to 5000 ns (0.1 MHz) with PerfectSquare
technology
Damping
50
Ω
, 400
Ω
Table 8 Receiver
Parameter
Specifications
Gain
0 db to 110 dB
Maximum input signal
20 Vpk
Receiver input impedance
400
Ω
±5 %
Receiver bandwidth
DC to 26.5 MHz at
−
3 dB (Standard version)
0.2 MHz to 26.5 MHz at
−
3 dB (EN12668 compliant version)
Digital filter settings
8 digital filter sets (standard version)
7 digital filter sets (EN12668 compliant version)
Rectification
Full-wave, Positive Half-wave, Negative Half-wave, RF
System linearity
Horizontal: ±0.5 % FSW
Resolution
0.25 % FSH, amplifier accuracy ±1 dB
Reject
0 % to 85 % FSH in 1 % increment positions
Amplitude measurement
1.25 % to 110 % full screen height
Measurement rate
Equivalent to PRF in all modes (single shot)
Table 9 Calibration
Parameter
Specifications
Automated calibration)
Velocity, Zero Offset
Straight Beam (first back wall or echo-to-echo)
Angle Beam (Soundpath or Depth)
Test modes
Pulse Echo, Dual, or Through Transmission
Содержание EPOCH 6LT
Страница 6: ...DMTA 10083 01EN Rev 3 June 2017 Table of Contents vi...
Страница 8: ...DMTA 10083 01EN Rev 3 June 2017 List of Abbreviations viii...
Страница 24: ...DMTA 10083 01EN Rev 3 June 2017 Important Information Please Read Before Use 16...
Страница 26: ...DMTA 10083 01EN Rev 3 June 2017 Introduction 18...
Страница 42: ...DMTA 10083 01EN Rev 3 June 2017 Chapter 2 34...
Страница 52: ...DMTA 10083 01EN Rev 3 June 2017 Chapter 4 44...
Страница 60: ...DMTA 10083 01EN Rev 3 June 2017 List of Figures 52...
Страница 62: ...DMTA 10083 01EN Rev 3 June 2017 List of Tables 54...
Страница 66: ...DMTA 10083 01EN Rev 3 June 2017 Index 58...