DMTA-10055-01EN, Rev. A, February 2015
Calibration
177
4.
Use the G1Start button (P5 key) to position gate 1 so that the reflection from the
first side-drilled hole is exceeding the gate threshold. This reflection should be
close to 12.5 mm (0.5 in.) (see Figure 9-26 on page 177).
Figure 9-26 Example of a gated signal for zero calibration
5.
Press dB, and then adjust the gain so that the echo amplitude is at approximately
80 %.
The AUTO XX% feature can be used to automatically adjust the gain to set the gated
echo amplitude to XX% of full screen height (the default XX value is 80 %). To activate
this feature, press 2
ND
F, (AUTO XX%).
A thickness measurement reading appears in large text above the A-scan.
6.
Once a steady reading is achieved, choose
Auto Cal > Cal-Zero
.
The screen freezes and the
Cal-Zero
box appears (see Figure 9-27 on page 178).
Thickness measurement
Содержание EPOCH 650
Страница 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Страница 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Страница 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Страница 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Страница 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Страница 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Страница 153: ...DMTA 10055 01EN Rev A February 2015 Programmable Inputs and Outputs 141 Figure 8 1 The A Out setup page...
Страница 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Страница 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Страница 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Страница 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Страница 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Страница 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...