Figure 9-27 Entering the Zero Cal thickness value
7.
Adjust the value to match the known thickness of the gated indication (0.500 in. in
this example), and then choose
Continue
to continue to the second calibration
step (see Figure 9-28 on page 179).
If for any reason you need to exit without acquiring calibration data, press
Cancel
.