EMP-H35S, EMP-H65S & EMP-H100S QUICK REFERENCE GUIDE
5
L A S E R O P E R A T I O N
Connect the laser fiber
H35 Standby and Ready screens
H65 & H100 Standby and Ready screens
n
The fiber is connected to the device
through the fiber port on the front panel.
n
The fiber connector has an additional
ring which facilitates the clamping of
the fiber to the connector on the device.
Furthermore, this ring enables the
automatic detection of fiber status
(present/absent), its diameter and its
type (single use/reusable). If the fiber
is not connected to the device, an error
message is displayed when the device
is switched on.
n
On the H65 and H100, a dedicated sensor
recognizes the presence of the user
hand, opening the external protection
shutter of the fiber connector.
Standby/Ready
n
When the user presses the Ready/Standby buttons, the system will be set in the Ready or Standby mode,
respectively. The button on the footswitch, or the Ready/Standby display area can be used interchangeably.
n
In Standby mode, the laser is not firing and the system cannot emit energy.
n
In the Ready mode, the laser is firing and ready to emit energy.